63
13.4.2
/Fault_out
Pin 16 (/Fault_out) indicates when a fault occurs or the Instrument is in protection status.
During normal operation, the voltage level of this pin stays high (5 V).
The voltage level will be low (0 V) if any of the circumstances below occurs:
4.
Interface CAN Error
5.
Module CAN Error
6.
SW Over Current Prot
7.
SW Over Voltage Prot
8.
SW Over Power Limit
9.
SW Over Max VA Limit
10.
Remote Inhibit ON
13.4.3
/Transient
When the output status changes, pin 21 (/Transient ) will send out a 500 uS, 5 V to 0 V
pulse to synchronize with another external device as shown in the below timing diagram.
This function is not available in the Program modes.
To active this function, follow the settings: Configure → Config 2 → Transient → ON.
Figure 53 - Transient Timing Diagram
Содержание 9830B Series
Страница 1: ...1 9830B Series USER MANUAL Model 9832B 9833B High Power Programmable AC Power Sources ...
Страница 10: ...ix 2 Notations TEXT Denotes a softkey TEXT Denotes a front panel key ...
Страница 28: ...15 6 Menu Tree ...
Страница 86: ...73 THD 11 Harmonic 3 weight 21 3 THD 12 Harmonic 3 weight 24 5 ...
Страница 87: ...74 THD 13 Harmonic 2 5 7 8 weight 2 3 9 8 15 8 2 5 THD 14 Harmonic 2 5 7 8 weight 1 15 4 9 7 9 1 25 ...
Страница 88: ...75 THD 15 Harmonic 5 7 weight 2 45 3 95 THD 16 Harmonic 3 5 7 9 weight 11 4 05 2 1 3 ...
Страница 89: ...76 THD 17 Harmonic 3 5 9 weight 7 17 3 42 0 8 THD 18 Harmonic 3 5 9 weight 8 11 3 48 1 ...
Страница 90: ...77 THD 19 Harmonic 3 5 9 weight 9 38 3 44 1 15 THD 20 Harmonic 3 5 7 9 11 weight 2 1 8 1 6 1 23 0 9 ...
Страница 95: ...82 THD 29 Harmonic 3 5 7 9 11 13 15 weight 33 3 20 13 8 10 8 8 5 7 2 5 5 ...
Страница 103: ...90 1 The instrument will show Cal End then press ESC to complete the external voltage calibration ...
Страница 108: ...95 22820 Savi Ranch Parkway Yorba Linda CA 92887 www bkprecision com 2020 B K Precision Corp Printed in Taiwan v081720 ...