87
1.
Connect the DMM to both ends of the shunt. Set the DMM to measure AC voltage.
2.
Press the System
→
Next
→
Calibration
→
softkeys. Enter the password 13579.
3.
Then select Imeas (AC current) on the calibration menu.
Freq1, Freq2, Freq3 and Freq4 represent output frequency of 100Hz, 400Hz, 800Hz and
1200Hz respectively.
5.
Press Freq1 key then press the Cal. to start calibration, and the instrument will
output the Low Point current.
Divide the voltage shown on the DMM by 3 (3
Ω
shunt) and use the number keys to enter
the result. Press the ENTER key when complete.
The instrument will show “OK” after
Low Point(A).
6.
Press Cal.
softkey to start calibration 2, and the instrument will output the ML Point
current.
Divide the voltage shown on the DMM by 3 (3
Ω
shunt) and use the number keys to enter
the result. Press the ENTER key when complete. The instrument will show “OK” after the
ML Point(A).
7.
Press Cal.
softkey to start calibration 3, and the instrument will output the MH Point
current.
Divide the voltage shown on the DMM by 3 (3
Ω
shunt) and use the number keys to enter
the result. Press the ENTER key when complete. The instrument will show “OK” after the
MH Point(A).
8.
Press Cal.
Softkey to start calibration 4, and the instrument will output the High
Point current.
Divide the voltage shown on the DMM by 3 (3
Ω
shunt) and use the number keys to enter
the result. Press the ENTER key when complete.
The instrument will show “OK” after the
High Point(A). The calibration date and time will be shown below.
Next, calibrate Freq2, Freq3 and Freq4 in order to complete the current calibration.
15.1.5
Restore to Factory Default (RECALL DATA)
If you do not have a device to calibrate the instrument, this step will restore the
instrument to the factory defaults.
In Setting and Measurement screen, press System
→
Next
→
Calibration
→
softkeys.
Enter the password 13579
→
RECALL DATA softkey. Use
◄
►
keys (or Rotary) to select
YES, then press ENTER to confirm.
Содержание 9830B Series
Страница 1: ...1 9830B Series USER MANUAL Model 9832B 9833B High Power Programmable AC Power Sources ...
Страница 10: ...ix 2 Notations TEXT Denotes a softkey TEXT Denotes a front panel key ...
Страница 28: ...15 6 Menu Tree ...
Страница 86: ...73 THD 11 Harmonic 3 weight 21 3 THD 12 Harmonic 3 weight 24 5 ...
Страница 87: ...74 THD 13 Harmonic 2 5 7 8 weight 2 3 9 8 15 8 2 5 THD 14 Harmonic 2 5 7 8 weight 1 15 4 9 7 9 1 25 ...
Страница 88: ...75 THD 15 Harmonic 5 7 weight 2 45 3 95 THD 16 Harmonic 3 5 7 9 weight 11 4 05 2 1 3 ...
Страница 89: ...76 THD 17 Harmonic 3 5 9 weight 7 17 3 42 0 8 THD 18 Harmonic 3 5 9 weight 8 11 3 48 1 ...
Страница 90: ...77 THD 19 Harmonic 3 5 9 weight 9 38 3 44 1 15 THD 20 Harmonic 3 5 7 9 11 weight 2 1 8 1 6 1 23 0 9 ...
Страница 95: ...82 THD 29 Harmonic 3 5 7 9 11 13 15 weight 33 3 20 13 8 10 8 8 5 7 2 5 5 ...
Страница 103: ...90 1 The instrument will show Cal End then press ESC to complete the external voltage calibration ...
Страница 108: ...95 22820 Savi Ranch Parkway Yorba Linda CA 92887 www bkprecision com 2020 B K Precision Corp Printed in Taiwan v081720 ...