(Master – to be added to the documentation)

M688 70 01 05 F04 / 3891 7319 (GB)

Sheet 2 of 3

 

1 -

Technical Safety Check (TSC)

EN

5 -   2

  Perfusor® Space, 3.0

Operation

Syringe fastening

Syringe recognition

Infusion

Buttons on the operating unit

Staff call

Trigger bolus at the device

Trigger bolus by pressing the PCA 
button

PCA lock (bottom side of pump) with 
inserted 50/60 ml syringe

PCA lock Syringe Anti Removal CAP PSP 
with inserted 50/60 ml syringe

Switch on unit without power supply

Self-test

Magnetic function of the battery 
compartment cover

Pressure cut-off

Syringe type: “#LehreOPS50”
Delivery rate: 200 ml/h

- Strain gauge pressure measurement

Pressure stage 1 (9…15N)________N

Pressure stage 3 (25…33N)_______N

Pressure stage 8 (63…76N)_______N

Error message “Alarm / pressure too 
high” at every pressure stage

- Motor power limitation

Pressure stage 1 (11…28N)________N

Pressure stage 3 (30…49N)________N

Pressure stage 6 (58…80N)________N

Error message “Alarm / drive blocked” 
at every pressure stage

Visual Inspection

Electrical Safety

According to IEC / EN 60601-1 

or VDE 0750 and VDE 0751

Functional Inspection

Содержание Perfusor Space

Страница 1: ...nglish 0 12 0 This document is not a complete Service Manual It only encompasses the pages that describe how to determine the age and how to check the condition of batteries as well as the accordingly adapted template for Technical Safety Check ...

Страница 2: ...e the pump from use and return it to the Medical Engineering Department for replacement of the battery In this case the battery has to be replaced regardless of its age or any further test Disassembly Note Move the drive head to the extended end position before starting disassembly see General pg 3 1 1 Press the lock Fig 3 8 Item 2 on the battery compart ment cover Fig 3 8 Item 1 and remove the ba...

Страница 3: ...h blade Syringe fastening Membrane in drive head Axial clearance of drive Screw covers Connectors P2 and P3 Protection cap at drive head Accessories Cleanliness completeness damage and faults affecting safety damage and readability of the label Check the unit and the accessories for compatibility The patient and housing leakage current of the Perfusor Space is caused exclusively by the operating v...

Страница 4: ...itch on unit without power supply Self test Magnetic function of the battery compartment cover Pressure cut off Syringe type LehreOPS50 Delivery rate 200 ml h Strain gauge pressure measurement Pressure stage 1 9 15N ________N Pressure stage 3 25 33N _______N Pressure stage 8 63 76N _______N Error message Alarm pressure too high at every pressure stage Motor power limitation Pressure stage 1 11 28N...

Страница 5: ... connector SP __________________________ __________________________ Power Supply SP Battery module Staff call lead Space PCA Kit PCA button Syringe Anti Removal Cap PSP ___________________________ Test result Defects found which could endanger patients users or third parties Yes No Inspection performed by Measures to be taken Repair Replacement of battery recommended _________________________ Unit...

Страница 6: ...Functional Check 1 Determine age of battery Open battery compartment cover and remove battery Read the batch number printed on the type plate of the battery and determine the age of the battery a For NiMH batteries see Fig 6 2 b For Li Ion batteries see Fig 6 3 Note If the battery is older than 4 years recommend the replacement of the battery or perform the battery check Fig 6 2 NiMH batteries Fig...

Страница 7: ...units from SW J on LEDs yellow red blue light up for a short moment LED changes from red to green Audible alarm A deep and a high sound Status display Battery capacity mains operation service connector SP Lighting Illumination of syringe compartment LC display and but tons 4 Syringe fastening a Axial fastening holds syringe wing b Syringe holder blade fixes syringe piston c Both claws in drive hea...

Страница 8: ...d The red LED on the service connector SP lights up 9 PCA lock on the Perfusor Space a Check the PCA lock with an approved 50 60ml syringe 10 PCA lock of the Syringe Anti Removal Kit PSP a Place on the drive head and lock b Check the Syringe Anti Removal Kit PSP with an ap proved 50 60ml syringe c Unlock and remove from the drive head Battery Alarm 1 Switch device off 2 Pull off the power supply 3...

Страница 9: ...Revision Documentation 9 EN 9 2 Perfusor Space 3 0 Version 3 0 Changed TSC of the unit Revised procedural instructions on the TSC Release date 25 10 2013 ...

Страница 10: ...B Braun Melsungen AG Carl Braun Strasse 1 34209 Melsungen Germany Tel 49 0 5661 71 0 Fax 49 0 5661 71 4567 www bbraun de Mat No M688_001152 Doc No 8713 9020 Version 3 0 Effective 25 10 2013 ...

Отзывы: