Perfusor® Space, 3.0
5 - 1
EN
5 -
Technical Safety Check (TSC)
Perfusor® Space
Checklist for Technical Safety Check – Every 24 Months
Unit: Perfusor®Space
Manufacturer: B. Braun Melsungen AG
User
Observe the Service Manual and the instructions for use. All measured values are to be documented.
Accessories used should be included in testing. Make exclusive use of calibrated measuring equip-
ment.
Article No.
Unit No.
Year of Procurement
Stock No.
Visual Inspection
Electrical Safety
According to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Functional Inspection
❒
Perfusor® Space:
Cleanliness, completeness, damage and
defects affecting safety, damage and
readability of the labels.
Particularly:
❒
Syringe holder with blade
❒
Syringe fastening
❒
Membrane in drive head
❒
Axial clearance of drive
❒
Screw covers
❒
Connectors “P2” and “P3”
❒
Protection cap at drive head
❒
Accessories:
Cleanliness, completeness, damage and
faults affecting safety, damage and
readability of the label
❒
Check the unit and the accessories for
compatibility
The patient and housing leakage current of
the Perfusor® Space is caused exclusively by
the operating voltage supply (Power Supply
SP or SpaceStation).
The Technical Safety Checks of the power
supply SP (drawing No. M001321005 F04)
or of the SpaceStation (drawing No.
M690000046 F04) serve to check whether
both limit values are met.
❒
Locking with second unit
❒
Operating unit magnets
❒
Battery compartment cover
Determine age of battery
❒
If older than 4 years:
recommend replacement of battery
or perform battery check
(refer to
➨
“Procedural Instructions
on the TSC“
)
Switch on unit with power supply
❒
Self-test
❒
Indicator lamps (LEDs)
❒
Audible alarm
❒
Visual alarm
❒
Status display
❒
Lighting
❒
Battery alarm
5 - 4