78
Error messages and troubleshooting
Error message
Cause
What to do
The reading is above 600 mg/dL.
In case of doubt with regard to the blood glucose level, check
the meter with the control solution. If the result is normal, re-test
with a blood sample 2 to 3 times. If the
Hi
(high) message persists,
please consult a physician promptly for assistance.
The reading is below 10 mg/dL.
In case of doubt with regard to the blood glucose level, check
the meter with the control solution. If the result is normal, re-test
with a blood sample 2 to 3 times. If the
Lo
(low) message persists,
consult a physician promptly for assistance.
Low battery voltage indicator.
The battery symbol is displayed during use.
Change the batteries immediately.
The battery voltage is no longer sufficient to turn
on the Omnitest® 5. The battery symbol flashes if a
button is pushed or if a test strip is inserted.
Change the batteries immediately.