Error messages and troubleshooting
Error message
Cause
What to do
There is a problem with the meter.
An incorrect signal from the check strip occurred.
Remove both batteries and insert them again. Repeat the test with
the check strip. If the error message continues to be displayed, the
meter must be replaced.
The error message may have been caused by a
used or wet test strip.
Repeat the test with a new test strip.
There was not enough blood or control solution
applied.
Repeat the test with a new test strip and sufficient volume of
blood or control solution.
There is a problem with the test strip.
The test strip is damaged. Repeat the test with a new test strip.
The blood sample or control solution was applied
before the symbol
was shown on the display.
Repeat the test with a new test strip. Apply blood or control
solution only after the symbol
has been shown on the display.