Guide for Short Infusion Pole
VOL
Infusion Volume
Press key below
VOL
. Enter volume
(0.1 ... 9999.9 ml). Confirm. Press again key
below
VOL
.
TIME
Infusion Time
Press key below
TIME
. Enter time e.g. 50 min as
5 0 or 2 h 30 min as 2 3 0. Confirm. Press again
key below
TIME
.
RATE
Delivery Rate
Only active when rate is calculated automatical-
ly. The key below
RATE
confirms the calculated
rate.
SF
Special Functions
If activated: dose calculation / bolus function /
standby / drug selection / occlusion pressure /
drop control / piggyback / battery capacity /
data lock / contrast / date, time.
Function Keys
Mains/Power Connection
(protect against ingress of moisture). In case of
mains/power failure, the pump switches to
battery operation. Battery operation time:
> 3.5 h at highest delivery rate. Automatic over-
load protection. Mains fuse: directly above the
mains/power connector.
Potential Equalisation
To be connected for
CF-applications.
Aluminium Housing
Easy-care, drip water protected, resistant to
disinfectants.
Mains/Power Supply
For operation with fluid manager system.
Multi-Function-
Connector (MFC)
Connection for staff
call, ambulances
(12 V DC) and inter-
face for fm anaes-
thesia/fm intensive.
Optical Interface
Infrared Interface for operation on
"fluid manager system”.
Flow Inhibitor
Opening door clamps off infusion line
automatically.
Peristaltic Pump
For precise and reliable dosage.
Door Opener
Special Functions "SF”
Function Keys
(Soft keys)
Mains/Power Switch ON/OFF
Suppress alarm tone for 2 min
Infusion START/STOP