ILLUSTRATION AND NOMENCLATURE
Figure 1
1.
E-Clip
2.
Fill Port
3.
Easypump C-bloc RA Pump
4.
Clamp
5.
Filter
6.
Flow Controller (tape to skin)
PRIMING THE ADMINISTRATION SET
• Open clamp.
• Remove tubing cap to prime pump (up to 15
minutes). When tubing is fully primed, close
clamp and replace tubing cap until ready for
use.
STARTING THE INFUSION
• Connect catheter to pump tubing. Open
clamp to begin infusion.
• Tape Flow Controller to skin to ensure flow
rate accuracy. Avoid contact with cold
therapy. Do not tape over filter.
CAUTION
• Temperature: The Easypump C-bloc RA Flow
Controller should be in direct contact with
the skin (88˚F/31˚C). Temperature will affect
solution viscosity, resulting in faster or slower
flow rate. If Easypump C-bloc RA is used
with the Flow Controller at room temperature
(68˚F/20˚C), flow rate may decrease by
approximately 25%.
See Table 1 on back for Flow Rate Accuracy
and Delivery Time information.