8.
Maintenance
The display "Battery replacement recommended" on the Acculan 3Ti char-
ger GA677 (see TA022160) indicates when the battery should be replaced
due to reduced capacity.
Note
The "Battery replacement recommended" display on the Acculan 4 charger
GA320 has no function in the NiMH battery GA666 and GA676.
No other maintenance measures must be performed on the battery.
For corresponding services, contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting
►
Replace product: have defective products repaired by Aesculap Technical Service, see Technical Service.
10. Storage and transport
►
In order to prevent damage during transport, ship product in the orig-
inal packaging and in an outer carton.
►
If the original packaging is no longer available, ensure the following
when packaging the product:
– The battery contacts are protected against short-circuit.
– The packaging completely encloses the product.
– The product is secured against movement within the packaging.
►
Completely charge the product upon receipt and before extended peri-
ods of storage.
11. Accessories/Replacement Parts
12. Technical Service
Modifications carried out on medical technical equipment may result in
loss of warranty rights and applicable licenses.
►
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Issue
Detection
Cause
Remedy
Battery not working
No signal tone when inserting the battery into
the machine
Battery defective
Replace product.
Battery quickly empty
Symbol "Battery replacement recommended"
illuminates at the end of the charging process on
the Acculan 3Ti charger GA677 (not on
Acculan 4 charger GA320)
Battery cells at end of life
Have battery cells replaced by
Aesculap Technical Service.
WARNING
Risk of injury from burns and damage to product in
case of short-circuit!
►
When storing the product, ensure that the prod-
uct cannot short-circuit with other products or
metal objects.
Art. no.
Designation
TA015514
Instructions for use, battery GA666, GA676 (A4)
TA011867
Instructions for use, battery GA666, GA676 (flyer)
WARNING
Risk of injury and/or malfunction!
►
Do not modify the product.