4.3.2
Sensing Head Cable
On all DM models (
exception 3M, P3, P7
) the sensing head cable can be shortened if necessary.
On the models
1M, 2M
and
DM Fast
the sensing head cable can be shortened by max. 3 m.
A shortening of the cable will cause an additional measuring error of about
0,1 K/ m
.
The
3M
models are only available with 3 m cable.
After you have modified the head code a reset is necessary to activate the change. [►6 Operating]
On the DM models LT21, LT21H and LT101H the head cable must not be moved during the measurement
You will find the calibration code on a label fixed on the head cable (near the electronics). Please do not remove this label or make
sure the code is noted anywhere. The code is needed if the electronics must be exchanged or in case of a necessary recalibration of
the sensor.