21
Setting up the Clearway in Timed Programmable Auto mode.
•
Switch on the clearway and select Prog. Timed Auto mode from the Mode select
list. (If Prog. Timed Auto does not appear in the list, this function is not available).
•
Press INS to select the insufflation pressure and use the
◄
- and +
►
buttons to
adjust to the required pressure.
•
Press Ti to select the insufflation time and use the
◄
- and +
►
buttons to adjust
to the required time.
•
Press Pause to select the pause time between insufflations and use the
◄
- and
+
►
buttons to adjust to the required time.
•
Press Ins Rpt. to select the number of insufflations before exsufflation. Use the
◄
- and +
►
buttons to adjust to the required number.
•
Press EXS to select the exsufflation pressure and use the
◄
- and +
►
buttons to
adjust to the required pressure.
•
Press Te to select the exsufflation time and use the
◄
- and +
►
buttons to adjust
to the required time.
•
Press Cycle Rpt. To select the number of insufflation/exsufflation cycles in the
sequence. Use the
◄
- and +
►
buttons to adjust to the required number.
•
Press menu and select treatment review screen to view the waveform if desired.
Press Menu again to escape.
•
Press Start/Stop to begin treatment.