20
Timed Programmable Auto mode
The Nippy Clearway can be pre programmed to deliver a timed sequence, which
could include one or more insufflations before exsufflation. This sequence can be
repeated up to 10 times if desired.
As an option, at the end of the sequence, two insufflations may be added to help the
patient to recover from the effort required during treatment. This is set through
advanced settings, see page 25.
For example, the sequence might be:-
2 x insufflation followed by exsufflation,
repeated twice.
In this example a 1 second insufflation @
10cm is followed by a pause of 2 seconds.
This is repeated and followed by a 1 second
exsufflation @ -10cm. The whole cycle is then
repeated.
Press the
Start/Stop
button to initiate
treatment. The clearway will start the
sequence and the display will change to the
active treatment display. This is a graphical
representation of the treatment sequence.
Insufflation is shown as a positive waveform
and exsufflation is shown as negative.
The
numbers in the top right hand corner of the
display indicate the progress. In this example
it is showing repetition number 1 of a total of 2.
Use the
Set
button to toggle between the
active treatment display and the normal
display if desired.
To aid synchronisation with the exsufflation, an audible beep will sound just before
the start of each exsufflation.
Press the
Start/Stop
button to end treatment.
A double beep will sound at the end of the sequence.
Manual override may be used at any time during the treatment and is available
for 1 minute after the treatment has finished to provide safer airway clearance.