
25
Extraoral Suction System, Instructions For Use and Installation
Statement of Manufacturer
—
Electromagnetic Immunity
Extraoral Suction System is designed for electromagnetic environment described below. Be sure to apply.
Electromagnetic
Immunity Test
IEC6061 Test Level
Test Level Conformance
Electromagnetic Environment
3V(
Effective Value
)
3V/m
The isolation distance between
portable and mobile RF
communications devices and any
part of the Extraoral Dentistry
Suction System, including cables,
should not be less than
recommended isolation distance.
The recommended isolation distance
is calculated by a formula
corresponding to the frequency
ofthe transmitter. Recommended
isolation distance formula:
RadioFrequency
Conduction GB/T
17626.2 Radio
3V(Effective Value)
150kHz
~
80MHz
d
=
d
=
80 MH
Z
~
800 MH
Z
d
=
800 MH
Z
~
2.5 GH
Z
P—based on the transmitter's
maximum rated output power
provided by
transmittermanufacturer, in watts
(W);
Frequency Radiation
GB/T 17626.3
3V/m80MHz
~
5GHz
d—Recommended isolation distance,
in meters (m).
The field strength of the fixed RF
transmitter is determined by
surveying the electromagnetic field
a, and in each frequency range
dshould be lower than the
compliance level. Interference may
occur near the equipment marked
with the following symbol.
Note 1: At 80MHz and 800MHz frequencies, apply higher frequency band formula.Note 2: These guidelines may
not be suitable for all situations. Electromagnetic propagation is affected by absorption and reflection from
buildings, objects and human bodies.
The field strengths of fixed transmitters, such as: base stations for wireless (cellular/cordless) phones and
terrestrial mobile radios, amateur radios, AM and FM radio broadcasts, and television broadcasts, cannot be
accurately predicted theoretically. To assess the electromagnetic environment of fixed RF transmitters, surveys of
electromagnetic sites should be considered. If the measured field strength of the Extraoral Suction System is higher
than the applicable RF compliance level above, the Extraoral Suction System should be observed to verify that it
can operate normally. If abnormal performance is observed, supplementary measures may be necessary, such as
reorienting or repositioning the Extraoral Suction System.In the entire frequency range of 150kHz to 80MHz, the
field strength should be lower than 3V/ m.
P
2
.
1
P
2
.
1
P
2
.
1
Содержание Optima EOS350
Страница 9: ...6 Extraoral Suction System Instructions For Use and Installation 4 1 Product Dimensions inch ...
Страница 10: ...7 Extraoral Suction System Instructions For Use and Installation ...
Страница 14: ...Extraoral Suction System Instructions For Use and Installation ...
Страница 40: ...Système d aspiration extra orale mode d emploi et installation 37 4 1 Dimensions de l appareil pouces ...
Страница 41: ...Système d aspiration extra orale mode d emploi et installation 38 ...
Страница 45: ...Système d aspiration extra orale mode d emploi et installation 42 ...
Страница 72: ...Extraorales Absaugsystem Gebrauchsanweisung und Installation 69 4 1 Produktabmessungen Zoll ...
Страница 73: ...Extraorales Absaugsystem Gebrauchsanweisung und Installation 70 ...
Страница 77: ...Extraorales Absaugsystem Gebrauchsanweisung und Installation 74 ...
Страница 105: ...Sistema de succión extraoral Instrucciones de uso e instalación 102 4 1 Dimensiones del producto pulgadas ...
Страница 106: ...Sistema de succión extraoral Instrucciones de uso e instalación 103 ...
Страница 110: ...Sistema de succión extraoral Instrucciones de uso e instalación 107 ...
Страница 137: ...Sistema di aspirazione extraorale Istruzioni per l uso e l installazione 134 4 1 Dimensioni del prodotto in pollici ...
Страница 138: ...Sistema di aspirazione extraorale Istruzioni per l uso e l installazione 135 ...
Страница 142: ...Sistema di aspirazione extraorale Istruzioni per l uso e l installazione 139 ...
Страница 169: ...Extraoraal afzuigsysteem Gebruiks en installatieaanwijzingen 166 4 1 Afmetingen van het product inch ...
Страница 170: ...Extraoraal afzuigsysteem Gebruiks en installatieaanwijzingen 167 ...
Страница 174: ...Extraoraal afzuigsysteem Gebruiks en installatieaanwijzingen 171 ...
Страница 200: ...System odsysania wydzieliny z jamy ustnej instrukcja obsługi i instalacji 197 4 1 Wymiary produktu cale ...
Страница 201: ...System odsysania wydzieliny z jamy ustnej instrukcja obsługi i instalacji 198 ...
Страница 205: ...System odsysania wydzieliny z jamy ustnej instrukcja obsługi i instalacji 202 ...