Symptom
Solution
The LEDs of the power supply do not
illuminate when the unit is switched
on
•
Check that the mains cord of the power supply is properly connected to the mains supply.
•
Check the power supply of the dental office.
•
•
Replace the transformer.
•
Send the unit and the accessories to an approved
BA
repair center.
International
No irrigation flow when the foot pedal
is pressed
•
Turn the liquid flow rate knob to the maximum.
•
Make sure that water comes in external device for the
OPTIMA
BAC100.
•
•
Use compressed air (max. 3 bar) and check that air comes out of the handpiece.
•
Press the bottle by hand and push the air through to tube to change the negative pressure status in whole
•
Replace the pump head for the
OPTIMA BAC200.
•
Check the connection of the foot pedal.
•
Replace the foot pedal.
•
Send the unit and the accessories to an approved
BA
repair center.
International
No ultrasound oscillations when the
foot pedal is pressed
•
Check that the handpiece is connected to the handpiece cord.
•
•
Check that the instrument is not cracked.
•
Check the connection of the foot pedal.
•
Replace the foot pedal.
•
Send the unit and the accessories to an approved
BA
repair center.
International
Ultrasound power decreasing or
inadequate
•
Check that the selector is beyond the position 4.
•
•
Check the wear of the instrument and replace if necessary.
•
Replace the handpiece.
•
Send the unit and the accessories to an approved
BA
repair center.
International
Water leaks between the handpiece
cord and the handpiece
•
Replace the O-ring on the handpiece cord.
•
Replace the handpiece.
•
Send the handpiece to an approved
BA
repair center.
International
•
Check the O-ring of the bottle cap.
•
Replace the bottle.
•
Send the unit and the accessories to an approved
BA International
repair center.
Insufficient lighting of the handpiece
•
Send the handpiece to an approved BA International repair center.
TROUBLESHOOTING
69
Turn the power
dial
of the unit to ON and check if the LED is lit.
Disconnect the
piezo
handpiece and check that the liquid exits from the cord.
C
heck
that the instrument is tight.
C
heck
that the instrument is tight.
•
Make sure that the bottle is properly installed for the
OPTIMA
BAC200.
•
Replace the water filter for the
OPTIMA BAC100.
Only
OPTIMA
BAC200: