23
Approved sterilization procedures
Follow your country-specific directives, standards and guidelines.
> Steam sterilization class B with sterilizers in accordance with EN 13060.
Sterilization holding time minimum of 3 minutes at 135 °C.
or
> Steam sterilization class S with sterilizers in accordance with EN 13060.
The sterilizer manufacturer must expressly approve the sterilization of contra-angle handpieces.
Sterilization holding time minimum of 3 minutes at 135 °C.