7 Conditioning
Operating Instructions BA ULTIMATE
34
65 56 067 D3560
.
201.09.03.02
7.2
Automated cleaning and disinfecting ...
NOTICE! Do not carry out automated cleaning of the quick couplings.
7.2.1
... with cleaning and disinfection equipment
BA 695 / BA 688 can also be cleaned and disinfected in suitable
cleaning and disinfection equipment.
The cleaning and disinfection equipment used must be approved by its
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ BA 695 / BA 688 is conditioned with a cleaning and disinfection
device.
1. Check whether BA 695 / BA 688 is clean after conditioning under
good lighting (min. 500 lux) and color rendering index (min. 80 Ra).