Operating Instructions BA ULTIMATE
5 Operation
65 56 067 D3560
.
201.09.03.02
23
5
Operation
NOTICE! Use only burs and diamond polishers that are sharp and
undamaged. Use clean burs and diamond polishers to avoid dirt in the
clamping system.
CAUTION! A loose or partially removed bur instrument can detach
itself from the head or break off. This may cause injury! Therefore, use
the turbine only if the bur instrument is inserted at least 10 mm and is
clamped securely in place.
CAUTION! Insufficient cooling leads to overheating of the
preparation site and damage to the tooth substance. Ensure that the
water content is > 50ml/min.
CAUTION! Do not pull the patient's cheek back with the handpiece
while the turbine is rotating. This would actuate the pushbutton, thus
creating a risk of burning the patient's oral mucosa.
IMPORTANT: For application details and operating data of the different
preparation tools, please refer to the information provided by the
manufacturer.