17
Fault
Possible cause
Solutions
Light intensity is
weak.
1. The optical fiber is
not installed well.
2. There is crevice on
the optical fiber.
3. There is resin on the
tip of the optical fiber.
1. Reinstall the optical
fiber.
2. Change a new optical
fiber.
3. Clear the resin.
The equipment
is not charging
when the adapter is
connected.
1. The adapter is not
connected well.
2. Fault of adapter or
incompatible.
3. The charging point is
impurity.
1. Reconnect.
2. Change the adapter.
3. Cleaned by the alcohol.
Effective duration of
the battery become
short.
The capacity of the
battery decreased.
Change a new battery.
The mode indicator
twinkles when
charging.
1. Low voltage.
2. Short-circuit of the
battery.
1. Back to normal after
15 minuets charging.
2. Change a new battery.
If such handlings are completed, the equipment still cannot work
normally, please contact B.A. International or an approved dealer.
11 After service
The warranty of BA Optima 10 is one year from date of purchase. For
any technical issues within or outside the warranty period, please contact
B.A. International or an approved B.A. International dealer for repair.