
16
9 Maintenance
9.1 This equipment does not include self-maintenance parts. Please send
the unit to B.A. International or an approved dealer for repair.
9.2 Only the optical fiber of this equipment can be autoclaved under high
temperature and high pressure, other parts should be cleaned by clean
water or neutral sterilized liquid, but do not soak the equipment in the
water. Do not clean by volatile or soluble liquid, otherwise the marks of
the control panel will fade.
9.3 Please clean the optical fiber to avoid the remaining resin on the
surface and infect the life-span and the effectiveness of solidification.
10 Troubleshooting
Fault
Possible cause
Solutions
No indication, no
response.
1. Battery is out of
power.
2. Fault of battery.
3. The main unit battery
protection system
works.
1. Charge the equipment/
Replace the battery.
2. Replace the battery.
3. Place the main unit
into the socket on the
charger for activation.
“Er” shown on the
screen.
Fault of main unit.
Send to after service for
repair.
Wink shown on the
screen.
Low battery.
Reconnect the charger, if
"Er" show again after 15
minutes please change
the battery.
Содержание BA-OPTIMA 10
Страница 158: ...156 ...
Страница 159: ...157 ...