background image

 

 
 
 
 

 

 
 
 
 
 
 

 
 
 

 

 

 

 

 
 

 

 
 

 
 
 
 

 
 
 

 

 

USER  
MANUAL 

 

 

AKA/G SERIES 

 

 
 
 

 
 
 
 

File: 2017-10-27 AKA_G AKAG122 GB 

 

Содержание AKA Series

Страница 1: ...USER MANUAL AKA G SERIES File 2017 10 27 AKA_G AKAG122 GB ...

Страница 2: ...t calibration options 19 15 3 Auto zeroing function 23 15 4 Unit selection 23 15 5 Interface parameters setting 25 15 6 Print setup 26 15 7 LCD settings 28 15 8 Language selection 28 15 9 Setting date and time 29 15 10 Keyboard options 29 15 11 Analog output 30 15 12 Speed 30 16 Applications 30 16 1 Product and user identification 32 16 2 Pieces counting 33 16 3 Unit 34 16 4 Percentage 35 16 5 Ani...

Страница 3: ...gal verification comply with certificate of type approval and are marked with the following legal and securing items green metrological mark placed on the balance name plate notified body stamp number of notified body on the balance name plate protective seals placed on an edge of balance name plate the casing mounting screw and in the place of access to adjustment switch In order to renew legal v...

Страница 4: ... type supplied with the balance and supply voltage have to be compatible with specified technical data Do not use the balance when its cover is opened Do not use the balance in explosive conditions Do not use the balance in high humidity environment If the balance seems not to operate properly switch it off and do not use until checked by authorised service According to current acts of low about p...

Страница 5: ...on weight OIML F2 100g F2 200g F2 200g F1 500g Type AKA1200G AKZ1200G AKA2200G AKZ2200G AKA3200G AKZ3200G AKA4200G AKZ4200G AKZ10 Capacity Max 1200g 2200g 3200g 4200g 10kg Min load Min 0 5g 0 5g 0 5g 0 5g 5g Reading unit d 0 01g 0 01g 0 01g 0 01g 0 1g Verification unit e 0 1g 0 1g 0 1g 0 1g 1g Tare range 1200g 2200g 3200g 4200g 10kg Accuracy class II Working temperature 18 33 C Weighing time 3s Pa...

Страница 6: ..._________________ 5 General balance description AKA100G AKA520G AKZ100G AKZ520G view 1 pan 2 pan support 3 pan ring 4 graphical display 5 keys 6 rotating legs 7 level 8 draft shield AKA1200G AKA4200G AKZ1200G AKZ10G 1 pan 2 pan support 4 graphical display 5 keys 6 rotating legs 7 level 7 6 1 2 4 3 5 8 7 6 1 2 4 5 ...

Страница 7: ..._________________________________________________________________________ Feeder 230V 50Hz Printer or computer ON OFF Adjustment switch RxD receiver TxD transmitter GND Case RS232C Feeder 1 2 3 4 5 6 7 8 9 Data 5V GND Clock PS2 1 2 3 4 5 6 Keyboard USB option ...

Страница 8: ...n go to option above navigation go to option below navigation enter into option navigation exit from option CLR undo last operation I switch on switch off standby indicator shows stabilization of weighing result linear indicator indicator of balance load 0 100 OFF indicator appears after the balance is switched off with I key decrease of last digit informs that reading unit value is lower than acc...

Страница 9: ...can interrupt working balance This location has to maintain proper temperature for working balance and necessary space for its operating The balance should stay on stable table made of material that does not influence magnetically on the balance Rapid air blasts vibrations dust rapid temperature changes or air humidity over 75 are not allowed in balance surrounding The balance should be far from h...

Страница 10: ...water level 7 at the back of the balance is in the middle 4 for AG100 AG600 Gently insert the mandrel of pan support 2 into balance mechanism socket through the pan ring 3 and the pan 1 on AG600 balances have not the pan ring 5 for AG1000 AG4000 Place nuts 2 on mandrels that are visible in balance cover holes put the pan 1 on nuts 6 If the balance was taken from a lower temperature surrounding to ...

Страница 11: ... e make sure that 0 zero indicator is displayed before sample is placed on the pan If not press 0 key and wait until the balance is zeroed and zero indicator appears In other balances the key does not operate 6 When the balance is not used but should be ready to work immediately it can be switched off by pressing I key The backlight of balance reading system is then switched off and the balance en...

Страница 12: ...tion is the process of putting internal weight on automatically by balance mechanism and correcting accuracy in balance firmware The correction is necessary because of differences between values of gravitational acceleration in the place where the balance was manufactured and in the place where it is operated as well as due to changes of balance level and temperature Internal calibration is perfor...

Страница 13: ...uses the balance being blocked Until calibration process is finished do not perform any operation on the balance Any vibrations and shocks interfere calibration process and may delay it or deteriorate accuracy of its result When internal calibration is performed successfully the balance indicates zero on the display at empty pan Note In order to terminate calibration process press CLR key and wait...

Страница 14: ...or additional charge is given in technical data table In order to calibrate the balance with external weight in legally verified balances verification seals should be removed and another legal verification should be performed In this case it is recommended to contact authorized service centre Calibration with external weight is described in details in chapter 17 1 13 Connecting the balance to comp...

Страница 15: ... If the scale is equipped with two serial joints interfaces Print and SErIAL function is set independently for both interfaces If scale cooperates with a computer then the computer must have a special program Dedicated programs are also offered by AXIS Needed drivers and instructions can be found on the CD supplied with Axis scales ...

Страница 16: ...tion choosing scrolling suboption ENTER enter option selection CLR exit actual option undo last operation auto calibration stop F1 F2 F5 fast access to application shortcut activated in application suboption MENU enter exit from menu turned on off active application To use option or to choose application move cursor and press ENTER key Important tool to navigate are fast access keys which are assi...

Страница 17: ... Keyboard keys options Analog out 4 20mA 0 10V out configuration option Speed weighing speed selection Firmware update firmware actualization only for service Defaults back to factory settings Exit Attention Using Defaults option doesn t change basic metrological balance parameters like sensitivity linearity if the calibration switch isn t moved but all other settings that have influence on balanc...

Страница 18: ...t after pressing MENU key choose Menu option List of applications will show up Precise description of all applications in Applications chapter Adding applications to personalized menu is done by pressing ENTER key when chosen application is highlighted Added application is marked with V sign After choosing all necessary applications use Exit option Defaults option deletes all applications from per...

Страница 19: ...gal verification To renew legal verification of the balance it is necessary to contact a service or notified body In balances comply with verification requirements performing calibration requires changing adjustment switch position which is placed behind protecting mark sticker of a notified body An access to the switch is possible only after removing the mark Before proceeding with calibration fo...

Страница 20: ...ing standard of mass value which will be used to calibrate it is suggested to use standard of mass value close to balance s max SETUP 1 Menu 2 Calibration 3 Auto zeroing 4 Unit 5 Interface 6 Print setup 7 LCD settings 8 Language 9 Time date 10 Keyboard 11 Analog output 12 Firmware update 13 Defaults 14 Exit CALIBRATION 1 Calibration 2 Mode 3 Load 4 Report 5 Linearity 6 Exit CALIBRATION 1 Calibrati...

Страница 21: ... means that calibration process ended Besides Report option all other calibration options are available after switch position change SETUP 1 Menu 2 Calibration 3 Auto zeroing 4 Unit 5 Interface 6 Print setup 7 LCD settings 8 Language 9 Time date 10 Keyboard 11 Analog output 12 Firmware update 13 Defaults 14 Exit CALIBRATION 1 Calibration 2 Mode 3 Load 100g 4 Report 5 Linearity 6 Exit CALIBRATION P...

Страница 22: ...andard of mass used during last calibration internal standard of mass weight registered during last calibtration difference between internal standard of mass factory actual CALIBRATION REPORT ALN220G MAX 220g e 0 001g d 0 0001g S N 1234 PROD DATE 2015 10 25 FIRM VER ALNG106 2015 10 23 AD7710 SIL FACTORY EXT LOAD 200 00 g FACTORY INT LOAD 196 131 g CALIBRATION NO 1 CALIBRATION DATE 2015 01 22 CALIB...

Страница 23: ...orrected automatically and when the pan is unloaded zero indication will be maintained regardless of environment conditions To turn on the function use navigation keys and ENTER key choose Status ON SETUP 1 Menu 2 Calibration 3 Auto zeroing 4 Unit 5 Interface 6 Print setup 7 LCD settings 8 Language 9 Time date 10 Keyboard 11 Analog output 12 Firmware update 13 Defaults 14 Exit AUTO ZEROING 1 Statu...

Страница 24: ...g pharmaceutical ounce Grain 1gr 0 06479891g grain Pennweight 1dwt 1 55517384g jewellery mass unit Gram 1g gram Readout for different units Unit Readout g 0 0001 g ct 0 0005 ct lb 0 0000001 lb oz 0 000001 oz ozt 0 000001 ozt gr 0 001 gr dwt 0 0001 dwt Unit selection is done with navigation keys and ENTER key SETUP 1 Menu 2 Calibration 3 Auto zeroing 4 Unit 5 Interface 6 Print setup 7 LCD settings ...

Страница 25: ...n network nr if the scale doesn t work in network the number must be 0 transmission through serial interface SendInG StAb transmission after key is used and result is stable noStAb transmission after key is pressed without need of stabilisation Auto automatic transmission after load is put on and result is stable Auto Cont continuous transmission about 10 results per second Cont Remove trans Defau...

Страница 26: ... user name Prn no successive printout number choose this option to zero counter Prod Id product number Prod bA product barcode inscribed or scanned Prod nA product name Count counting result PCS function APW unitary mass PCS function nEt net mass tArE current tare value GroSS gross mass totAL total mass totAL function During inscribing Prod Id barcode reader can be used optionally If the scale is ...

Страница 27: ...ormal weighing with clock option all printout positions deactivated Sample printout during normal weighing some printout positions activated 20 07 kg 20 04 kg 20 04 kg 20 07 kg 2012 11 08 10 01 20 04 kg 2012 11 08 10 01 20 04 kg 2012 11 08 10 01 AKA220G MAX 220g e 0 001g d 0 001g S N 30000017 USER ID 000001 DATE 2012 11 08 TIME 12 26 NO 3 PROD ID 01 COUNT 0 PCS APW 0 0000 g NET 213 8 g TARE 0 0000...

Страница 28: ...Language selection Function enables to set language Polish German Spanish English SETUP 1 Menu 2 Calibration 3 Auto zeroing 4 Unit 5 Interface 6 Print setup 7 LCD settings 8 Language 9 Time date 10 Keyboard 11 Analog output 12 Firmware update 13 Defaults 14 Exit LCD SETTINGS 1 Contrast 2 Backlight ON OFF 3 Negative ON OFF 4 Exit SETUP 1 Menu 2 Calibration 3 Auto zeroing 4 Unit 5 Interface 6 Print ...

Страница 29: ...ion enables to set options related to scale s keyboard sound when pressing keys and PS 2 interface activity SETUP 1 Menu 2 Calibration 3 Auto zeroing 4 Unit 5 Interface 6 Print setup 7 LCD settings 8 Language 9 Time date 10 Keyboard 11 Analog output 12 Firmware update 13 Defaults 14 Exit TIME DATE 1 Time 09 11 03 2 Date 2015 10 01 3 PIN 4 12 24 12H 24H 5 Form YYYY MM DD MM DD YYYY DD MM YYYY 6 Mai...

Страница 30: ...o environment conditions Options Default default weighing speed Slow slow speed measurement Medium medium speed measurement Fast fast speed measurement Attention When setting Fast option check if weighing results are stable If not use slower option SETUP 1 Menu 2 Calibration 3 Auto zeroing 4 Unit 5 Interface 6 Print setup 7 LCD settings 8 Language 9 Time date 10 Keyboard 11 Analog output 12 Firmwa...

Страница 31: ...hey don t conflict List of available applications Product ID assigning identification number to product User ID assigning identification number to user PCS pieces counting Unit actual weight unit selection Percent percentage conversion LabEL label number selection Animal animal weighing Tare setting storing tare value MAX MIN maximal value indication Newton indication in force units Total summing ...

Страница 32: ...printer and computer enables to build simple identification and archivisation system Inscribing multi digit data without using e g computer keyboard is inconvenient and using barcode reader is beneficial After selecting product and user it is possible to send to computer or printer actual scale indication with additional data selected by PrInt option SetuP Application options ID inscribing product...

Страница 33: ...W set unitary mass directly Shortcut fast access key selection F1 F2 or F5 Measurement consists of 3 phases Taring empty container or empty pan Single unit mass counting Counting pieces quantity in weighed portion Actions order Put empty container on pan and press T key Wait until indication is zero Put on a sample with pieces quantity earlier inscribed and press ENTER Balance shows pieces quantit...

Страница 34: ...les to choose actually used mass unit Chosen unit changes to default after balance turns off Selection of units Carat 1 ct 0 2 g Miligram 1mg 0 001g Kilogram 1kg 1000g Pound 1 lb 453 592374g Ounce 1oz 28 349523g Ounce troy 1ozt 31 1034763g pharmaceutical ounce Grain 1gr 0 06479891g grain Pennweight 1dwt 1 55517384g jewellery mass unit Gram 1g gram Readout for different units Unit Readout g 0 0001 ...

Страница 35: ...ases first phase weighing a reference sample 100 second phase measuring specific sample as a percentage of the reference sample Actions order Put empy container and press T Wait untill balance indication zeroing Put reference sample 100 and press ENTER Balance shows in percentage In order to end working with function press MENU key choose Percent and Deactivate APPLICATIONS Percent Exit PERCENT 1 ...

Страница 36: ...t empty box and press T Wait untill balance indication zeroing Put animal in box and press ENTER The balance makes a series of measurements and displays result After unloading the balance is ready for next measurement The balance will show stable averaged result and will send it through serial port To end working with the function press MENU key choose Animal and Deactivation APPLICATIONS Animal E...

Страница 37: ...hortcut fast access key selection F1 F2 or F5 Actions order Put an object on pan and press shortcut key chosen earlier to Tare setting application eg F2 Application options show up Choose Tare from scale or Tare value inscribe value and press ENTER and next Activation The balances shows indications with tare After pressing T or unloading pan the balance will show minus indications tare value subst...

Страница 38: ...dication Mode maximal value MAX or minimal MIN Shortcut fast access key selection F1 F2 or F5 Order of actions Perform series of weighing The balance indicates the max or min weight value T key causes balance zeroing and enables to start next series of weighing In order to finish working with the app press MENU key choose MAX MIN and Deactivation APPLICATIONS MAX MIN Exit MAX MIN 1 Activation 2 Mo...

Страница 39: ...options Activation force measurement start Shortcut fast access key selection F1 F2 or F5 Order of actions After activation the application is ready to work Exert force on pan do not overload the pan and the balance will show result in N unit In order to finish working with the app press MENU key choose Newton and Deactivation APPLICATIONS Newton Exit Newton 1 Activation 2 Shortcut F4 3 Exit Max M...

Страница 40: ...s Make series of weighings in Manual mode press key after each measurement In Auto mode measurement are stored automatically Each measurement inscribing is confirmed by displaying sum and average value Another recalling the application and option Preview or shortcut key use displays sum measurements quantity average value and available options register printout F3 reset zeroing memory CLR back to ...

Страница 41: ...Standard scale is set for cooperation with optical indicator On outputs P1 P3 Relays socket short circuit states appear as result of comparison scale indication with threshold values On the chart below output states are shown during increasing load on the scale for both working modes Indication mode Batching mode P3 P1 P2 P3 zero zero treshold treshold thr I thr I thr I thr II thr II thr II P1 P2 ...

Страница 42: ...on F1 F2 or F5 Actions order No load load smaller than zero threshold no signalization Put weight Case 1 Load under MIN threshold balance signalizes to small value MIN Case 2 Load above MIN threshold but under MAX balance signalizes good value OK in Impulse mode additional short sound signal occurs Case 3 Load above MAX balance signalizes too big value MAX in Impulse mode additional long sound sig...

Страница 43: ...dvised to use MS3K P electronic board sold separately consisting of RM96P transmitters with DC24V input voltage and AC250V 3A output Important notes 1 After switching the scale on both thresholds are set to maximum values 2 When setting upper threshold value pay attention that its value is not below lower threshold value 3 Setting lower and upper threshold value is possible after sending appropria...

Страница 44: ...is loaded and its indications stabilize After each loading printout is made with number of measurements result date and time if clock is installed and the function is activated For the obtained measurements series the scale evaluates n number of samples sum x sum of all samples n x x sum _ x average value sum x n min minimal value from n samples max maximal value from n samples max min maximal val...

Страница 45: ...uto mode measurements are saved automatically Each measurement inscribing is confirmed by displaying sum and average value Choosing app and Preview option or using shortcut key displays statistical results and available options stats register printout F1 displaying histogram F2 reset zeroing memory CLR back to summing In order to end Stats app choose the application choose Total and Deactivation A...

Страница 46: ...ove allowable upper value TOTAL sum of weights of all n samples AVERAGE average weight as Total n MIN minimum weight in n samples MAX maximum weight in n samples ST DEV standard deviation ST DEV standard deviation percentage Statistics function cooperation with computer and Printer Scale can be equipped with two serial ports marked as Port 1 computer and Port 2 printer After each data printout by ...

Страница 47: ...sts of two phases Phase I solid body sample measurement in the air Phase II measurement with immersion in the liquid This function also allows for determination of liquid density upon the basis of plunger weight with known density in the air and tested liquid The following formula is used m1 m2 ρ V where m1 plunger mass in the air m2 plunger mass in the liquid V plunger volume The plunger volume i...

Страница 48: ...y on upper tray measurement in air and press ENTER Put the examinated solid body on lower tray measurement in air and press ENTER Results are displayed and following options result printout CLR exit to summing In order to finish work with application choose application from menu and select Deactivation APPLICATIONS Density Exit DENSITY 1 Activation 2 Preview 3 Type of material Solid Liquid 4 Type ...

Страница 49: ...th T key Hang the plunger without dipping into liquid measurement in air and press ENTER Hang the plunger and dip it into liquid measurement in liquid and press ENTER Results are displayed and following options result printout CLR exit to summing In order to finish work with application choose application from menu and select Deactivation Max Min e d II AUT DEN T 0 000 g 0 100 Max Min e d II AUT D...

Страница 50: ...ut In order to print measurement results connect printer to scale s RS232C interface Way of connecting is described in chapter about scale communication After each measurement press to printout Printout sample Measurement number Mass in the air g Mass in the liquid g Density g cm3 Density with comp g cm3 Water density g cm3 Water temperature oC ...

Страница 51: ...ut vessel on pan and tare the balance using T key The balance is ready for weighing the successive ingredients and after each ingredient it is necessary to press T key It will zero the balance indications On the left side the sum of previously weighed ingredients and their number n is displayed In each moment in order to readout total mass use key second pressing causes return to weighing ingredie...

Страница 52: ...package weight Shortcut fast access key by navigation keys select F1 F2 or F5 F5 suggested Actions sequence Make series of weighing In manual mode press key after each measurement In Auto mode measurements are stored automaticallly if Threshold application is also active then to store measurement OK qualification is required Storing to memory is confirmed by displayed quantity of stored measuremen...

Страница 53: ...values This will allow to omit mistakes during big amounts of measurements and to connect goods control with selection full control When application is activated default threshold values are MIN Qn MAX QN 2T1 User can change the values Threshold app must be activated before goods control TP PREVIEW Defective 0 Disqualifying 0 Average 55 78 g Average qual 50 00 g Result POSITIVE printout F1 end F2 ...

Страница 54: ...E 20 TARE 5 00 g Qn 50 00 g Qn T1 45 50 g Qn 2 T1 41 00 g 1 55 76 g 2 55 77 g 3 55 77 g 4 56 15 g 5 56 16 g 6 56 15 g 7 56 15 g 8 56 15 g 9 56 16 g 10 56 16 g 11 56 15 g 12 56 15 g 13 56 17 g 14 56 84 g 15 56 15 g 16 56 31 g 17 56 16 g 18 56 59 g 19 51 81 g 20 51 81 g N 20 DEFECTIVE 0 DISQUALIF 0 MIN 51 81 g MAX 56 84 g AVERAGE 55 73 g AVER QUAL 49 15 g ST DEV 1 327 g RESULT POSITIVE SIGN ...

Страница 55: ...te 16 LF Attention Network number different than zero SErIAL nr function changes scale working mode communication with a computer is possible after logging the scale in with 02h scale number command To log the scale out use 03h command For example Using a program to test RS232 interface program is available on www axis pl in computer programs section for scale number 1 please write 0201 to log in ...

Страница 56: ... digits after dot should be the same as on scale display Scale Computer without response Example in order to set low threshold 1000g in scale B1 5 d 0 5g the following order should be sent S L 1 0 0 0 0 CR LF 53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah in order to set low threshold 100kg in scale B150 d 50g the following order should be sent S L 1 0 0 0 0 CR LF 53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah ...

Страница 57: ...r month day masa 10 signs scale indication mass unit P1 Steering instruction Attention 1 Except variable signs constant signs can also be inscribed e g factory name product name and so on 2 In standard only one label pattern is possible to printout number 0001 Using bigger amount of patterns other label numbers is possible thanks to LAbEL special function 3 To achieve label printout label printer ...

Страница 58: ...ssible cause Recommendation Test auto tests are in progress damage of electronic unit wait for 1 minute unfinished zeroing mechanical damage wait for 1 minute check if the balance is placed on stable ground not affected by vibrations Internal calibration load error too small load or overloading balance mechanism mechanical damage check if there are mounted all necessary pan elements or if there is...

Отзывы: