3. Set pulse width
The pulse width can be set from 30 µs to
400 µs. Press
control to enter this
setting. Then press the
or
control
to change the settings. If no instructions
regarding the pulse width are given in
therapy, set the control to 70 – 120 µs.
4. Set pulse rate
The pulse rate is adjustable from 2Hz
to 150Hz. Press
control to enter
this setting. Then press the “Increment”
or “Decrement” control to change the
settings. If no instructions regarding the
pulse width are given in therapy, set the
control to 70 – 120 µs.
5. Set timer
There are two adjustable timers available.
controls the application duration for the channels CH 1&2
controls the application duration for the channels CH 3&4
The settings are adjustable from 1 to 60 minutes. With the
setting
(C)
the machine runs in continuous mode. Press
control to enter the settings. Press
or
control to
adjust setting. Press the
control when the Timer shows
60 minutes to set the timer to continuous stimulation. Both
Timers can be adjusted in the same way.