Maximum voltage drop measured: Full range mA: 200mV
Maximum input current: 10A (not more than 10s);
Overload protection: 0.2A/250V fuse; 20A/250V fuse.
2.2.5. AC current measurement
Range / Accuracy / Resolution
2mA / ±(0.8%+5) / 0.1uA
20mA / ±(0.8%+5) / 1uA
200mA / ±(0.8%+5) / 10uA
20A / ±(2.0%+15) / 1mA
Maximum voltage drop measured: 200mV;
Maximum input current: 20A (not more than 15s);
Overload protection: 0.2A/250V fuse; 20A/250V fuse.
Frequency response: The frequency scope for all grades of current is 40Hz-1kHz; (applicable to standard
sine wave and triangle wave; and that for other waveforms: more than 200Hz, only for reference).
Display: Sine wave RMS (Average value response)
2.2.6. Resistance
Range / Accuracy / Resolution
200Ω / ±(0.5%+10) / 0.01Ω
2kΩ / ±(0.5%+8) / 0.1Ω
20kΩ / ±(0.5%+8) / 1Ω
200kΩ / ±(0.5%+8) / 10Ω
2MΩ / ±(0.5%+8) / 100Ω
20MΩ / ±(0.5%+8) / 1KΩ
200MΩ / ±(5,0%+10) / 10KΩ
Over load protection: 250V DC or AC peak value
At range 200Ω, short-circuit the test leads to measure the wire resistance, then subtracts it from the real
measurement.
2.2.7. Capacitance
Range / Accuracy / Resolution
20nF / ±(4,0%+10) / 1pF
200nF / ±(5,0%+20) / 100pF
2uF / ±(5,0%+20) / 100pF
20uF / ±(5,0%+20) / 1nF
200uF / ±(5,0%+20) / 10nF
2000uF / ±(5,0%+20) / 100nF
Overload protection: 36V DC or AC peak value