22
leads are shorten, but that will not affect the accuracy of measurement.
Overload protection: F 10A/1000V fuse for 10A range
F 0.63A/1000V fuse for
µ
A and mA ranges
Maximum input current:600mA dc or 600mA ac rms for
µ
A and mA ranges
10A dc or 10A ac rms for 10A ranges.
For measurements>7A, 4 minutes ON to measure 10 minutes OFF; Above 10A unspecified.
5. MAINETENANCE
This section provides basic maintenance information, including fuse and battery replacement
instructions.
Do not attempt to repair or service your Meter unless you are qualified to do so and have the
relevant calibration, performance test, and service information.
5.1.
General Maintenance
To avoid electrical shock or damage to the meter, do not get water inside the case. Remove
the test leads and any input signals before opening the case
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals:
Turn the meter off and remove all test leads.
Shake out any dirt that may be in the terminals.
Soak a new swab with a cleaning and oiling agent (such as WD-40)
Work the swab around in each terminal. The oiling agent insulates the terminals from moisture-
related contamination.
5.2.
Fuse replacement
Before replacing the fuse, disconnect test Leeds and/or any connectors from any circuit under
test. To prevent damage or injury replace the fuse only with specified ratings.
1)
Set rotary switch to the OFF position.
2)
Disconnect test leads and/or any connectors form the terminals.
3)
Use a screwdriver to unlock the four screws on the rear cover.
4)
Take out the rear cover from the meter.
5)
Remove the fuse by gently prying one end loose, then sliding that fuse out of its bracket.
6)
Install the replacement fuses only with specified ratings:
F 0.63A/1000V
∅
10.3
×
38 and
F 10A/1000V
∅
10.3
×
38