21
4.2.6.
Diode Test
Range
Resolution
Test Condition
2 V
0.001V
Forward DC current
approximately 1mA.
Reversed DC voltage
approximately 2.8V
4.2.7.
Continuity Check
Range
Resolution
Test Condition
660
Ω
0.1
Ω
Open circuit voltage: approx.
0.5V
Description: Continuity beeper
≤
30
Ω
4.2.8.
Capacitance
Range
Resolution
Accuracy
6.6nF
1pF
±(5.0% +5)
66nF
10pF
±(3.0% +3)
660nF
100pF
6.6
µ
F
1nF
66
µ
F
10nF
±(5.0% +3)
660
µ
F
100nF
6.6mF
1
µ
F
66mF
10
µ
F
Unspecified
4.2.9.
Current
Range
Resolutions
Accuracy
-55
°
~0
°C
0.1
°C
±(5.0% + 4
°C
)
1
°
~400
°C
±(2.0% + 3
°C
)
401
°
~1000
°
C
1
°C
±2.0%
Note: The specifications of temperature don’t include thermocouple errors.
4.2.10.
Current
DCA:
Range
Resolution
Accuracy
660
µ
A
0.1
µ
A
±(1.0% +3)
6600
µ
A
1
µ
A
66mA
0.01mA
±(1.5% +3)
600mA
0.1mA
10A
10mA
±(1.8% +5)
ACV:
Range
Resolution
Accuracy
660
µ
A
0.1
µ
A
±(1.5% +5)
6600
µ
A
1
µ
A
66mA
0.01mA
±(1.8% +8)
600mA
0.1mA
10A
10mA
±(2% +8)
Above accuracies can be guaranteed within 5%
~
100% of the full range.
The true RMS meter and the true RMS+PEAK meter has residual value within 10 counts when the test
Содержание AX-176