1.  Place the empty container or vessel on the weighing 

hook.

2.  Press the 

Z/T

 key once. The display will then reset 

to zero. If the container is removed, its weight will be 

displayed as a negative number.

3. 

You may now fill the container with the quantity to 

be weighed. The display will show the net weight of 

the contents.

4.  To return to normal weight display (gross weight), 

remove any items from the weighing hook and press 

the

 Z/T

 key again to re-zero the scale.

VIII.  Calibration

Initial calibration is performed at the factory. However, 

calibration may need to be checked and adjusted in some 

rare cases, or after heavy usage. To check calibration:

1.  Place the required calibration weight on the scale 

and note the displayed value.

2.  If the displayed value is not within acceptable 

tolerance (see chart below), adjustment is required.

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

PK-110

60g

20kg

OIML M2

IX. Adjustment

1.  Press and hold the 

Unit

 key until the display shows 

“CAL” followed by a random number (ISN).

2.  Make sure nothing is on the weighing hook then 

press the 

Z/T

 key.

3. 

The display will flash “20.00kg”. Place exactly 

Содержание PK-110

Страница 1: ...American Weigh Scales PK 110 0 44lb x 0 05lb 44 110lb x 0 1lb User Manual PK 110 Copyright 2012 American Weigh Scales Inc All rights reserved Rev 1 1...

Страница 2: ...ghing sensors and electronics causing inaccuracy or malfunction Avoid using your scale in close proximity to strong radio wave emitting devices such as cordless phones When weighing heavy items use pr...

Страница 3: ...urns the scale on or off 2 Zero Tare Key Re zeros the scale Used to remove the weight of a tray or container 3 Unit Key Tap to change the displayed unit of measure Hold to enter calibration see IX III...

Страница 4: ...s are properly aligned IV Operation 1 Press the ON OFF key to turn the scale on 2 Once the start up process completes and the scale goes to zero you are ready to weigh 3 Place the quantity to be weigh...

Страница 5: ...t when locked Hold Disabled Weight is no longer held on the display 3 Press the ON OFF key to confirm your selection and turn the scale off VI Changing the Unit of Measure 1 Press the Unit key to chan...

Страница 6: ...Calibration Initial calibration is performed at the factory However calibration may need to be checked and adjusted in some rare cases or after heavy usage To check calibration 1 Place the required c...

Страница 7: ...some cases nearby radio wave emitting devices can cause fluctuations Problem Displayed weight is inaccurate Solution Perform a calibration adjustment If you are still having trouble contact our warra...

Страница 8: ...lb Readability 0 05lb 0 1lb Min Weight 100g 0 2lb Power 2 x 1 5V AAA Alkaline Dimensions 4 8 x 3 5 x 1 0 in Hook Dimensions Gap 0 8 in Throat 1 0 in Warranty 10 Year Limited Warranty XIII Capacity Rea...

Отзывы: