background image

“CAL”. 

3. Press the [ M ] key again, and the display will show 

the required calibration weight (ie 500g or 200g). If you 

wish to use a 100g calibration weight instead, press 

the T key. The display will change to show 100g.

4. Place the calibration weight on the center of the 

platform and wait for the display to show “PASS”.

5. The scale will return to normal weighing mode. 

Calibration is complete!

IV. TROUBLESHOOTING

The primary reasons for inaccuracy or malfunction 

are low batteries, incorrect calibration, overload or 

operating on an unstable surface. Please keep this in 

mind and maintain and operate your scale properly. 

This scale is a precise instrument and must be 

handled with the utmost care and caution.

[Display]

  Elimination of errors

[88888]  

  Use the scale in a more stable position.

[EEEE] 

  You are overloading the scale. Remove the  

 

excess again after recalibration.

[Out2] 

 

The scale is out of calibration. Try   

 

recalibrating (see section III)