ACPro Series User Manual

Model

Acceptable 

Tolerance (±)

Calibration 

Weights

Recommended 

Accuracy Class

ACP-200

0.02g

100g, 200g

OIML M2

ACP-500

0.2g

500g

OIML M2

IX. Calibration Adjustment

1.  Press the 

ON/OFF

 key to turn the scale on.

2.  Press and hold the 

MODE

 key for three seconds 

until the display shows “CAL” then release.

3.  Press the 

MODE

 key again, and the LCD will flash 

the first calibration weight. Place the required weight 

on the middle of the platform

4.  The LCD will flash then flash the second required 

calibration weight. Place the required calibration 

weight on the middle of the platform. Skip this step 

for model ACP-500.

5.  The LCD will display “PASS” and return to normal 

weighing mode. Calibration is complete!

X.  Troubleshooting

•   

Problem:

 Scale will not turn on

•   

Solution: 

Check the battery

•   

Problem: 

Displayed weight fluctuates randomly

•   

Solution: 

Make sure your workspace is stable and    

 free of any air currents or vibration

•   

Problem:

 Displayed weight is inaccurate

•   

Solution:

 Perform a calibration adjustment

Содержание ACP-200

Страница 1: ...American Weigh Scales ACP 200 200x0 01g ACP 500 500x0 1g User Manual ACPro Series Copyright 2010 American Weigh Scales Inc All rights reserved Rev 2 0...

Страница 2: ...efore performing calibration if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avoid...

Страница 3: ...lation 1 Before using the scale for the first time check the battery compartment and remove any plastic insulation strips or packaging that may be in place to prevent battery drainage during shipping...

Страница 4: ...back on it will default to the last unit used ACP 200 Units g gram dwt pennyweight ct carat gn grain ACP 500 Units g gram oz ounce ozt troy ounce dwt pennyweight VI Using the Tare Feature You can use...

Страница 5: ...s the MODE key to change it between 25 50 75 or 100 and press PCS to confirm The count will be displayed on the bottom right of the LCD You may now add more items As items are remove items and the cou...

Страница 6: ...e the required weight on the middle of the platform 4 The LCD will flash then flash the second required calibration weight Place the required calibration weight on the middle of the platform Skip this...

Страница 7: ...m capacity exceeded Remove the extra weight to avoid damaging the load cell 3 LO Batteries are low Replace the batteries 4 UNSE The scale is unstable Use the scale in a more stable position ACP 200 AC...

Отзывы: