AwiLAB Gas analysis
9.1
Required Space
For information on the required space for the AwiLAB extension, please refer to the following
illustrations.
Figure 25: Required space for AwiLAB
9.2
Mechanical Connection
The reactors are 4/6mm connected to the milligas counters or the drum gas meters ( 4/6mmstainless
steel hose connections). From there, the 4/6mm hose lines lead into the AwiLAB via a condensate trap
AwiKon. Depending on the process, the discharge of the samples occurs either into a gas bag or via a
technically permanently leak-tight pipe, from the installation room normally into the open environment
(for more information on this, see
Chapter 5.2.6.4, "Exhaust air – Discharge of the Measuring Gas"
4/6mm line leads directly to the gas analysis system to measure the samples.
9.3
Electrical Connection
The milligas counters or the drum gas meters, respectively, are connected to the AwiLAB using a 3.5mm
jack plug (3-pin). The voltage supply for the AwiLAB takes place directly via the AwiFLEX gas analysis
system via a 3-pin (3 x 1.5²) feed line (230 V). The AwiLAB is controlled via RS485 bus. The assignment
can be found in the included electrical plan.