PROTECTING YOUR INSTRUMENT
TURN OFF INSTRUMENT WHEN NOT IN USE
The lifetime of the switching elements in the pulse generator module is proportional to
the running time of the instrument. For this reason the prime power to the instrument
should be turned off when the instrument is not in use. In the case of failure, the
switching elements are easily replaced following the procedure described in a following
section.
DO NOT EXCEED 10 kHz
The output stage may be damaged if triggered by an external signal at a pulse repetition
frequency greater than 10 kHz.
The maximum duty cycle is 0.5%.
USE A 50
Ω
LOAD
A 50 Ohm load is
required
for proper operation. The output stage may be damaged if
the output is not terminated into a 50Ω load.
22
Содержание AVRF-4A-B
Страница 34: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...
Страница 35: ...PCB 284B HIGH VOLTAGE DISCHARGE BOARD...
Страница 37: ...PCB 216E PRF PW TR ADJUST...
Страница 38: ...PCB 231A DC BIAS SOURCE...
Страница 44: ...PERFORMANCE CHECKSHEET 44...