background image

BASIC TEST ARRANGEMENT, WITH A REAL-TIME OSCILLOSCOPE

1. Connect the pulse generator to an oscilloscope using one of the two methods 

shown above. Note that:

a) If using a sampling oscilloscope, the use of 60 dB attenuator at the sampling 

scope vertical input channel will ensure a peak input signal to the sampling 
scope of less than 1 Volt.

b) If using a realtime oscilloscope, a 50 Ohm load must be used to terminate the 

output.

c) The SYNC output channel provides TTL level signals (approximately 0 and 

+3V). To avoid overdriving the TRIG input channel of some sampling 
oscilloscopes, a 20 dB attenuator should be placed at the input to the scope 
trigger channel.

d) The bandwidth capability of components and instruments used to display the 

pulse generator output signal (attenuators, cables, connectors, etc.) should 
exceed 2 GHz.

e) Set the oscilloscope to trigger externally with the vertical setting at 50 mV/div 

for sampling oscilloscopes, or 50 V/div for real-time scopes,  and the horizontal 
setting at 20 ns/div. 

2. Turn on the AVR-E3A-B. The main menu will appear on the LCD.

3. To set the AVR-E3A-B to trigger from the internal clock at a PRF of 10 kHz:

25

AVTECH
PULSER

MAIN OUTPUT

CONNECTOR

SYNC

OUTPUT

AC

POWER

ALL CABLES: 50 OHM COAXIAL

REAL-TIME

OSCILLOSCOPE

CHANNEL A

TRIG

INPUT

SCOPE
PROBE

TEST

LOAD

50 Ω

Содержание AVR-E3A-B

Страница 1: ...ND RS 232 CONTROL SERIAL NUMBER 14114 A V T E C H E L E C T R O S Y S T E M S L T D N A N O S E C O N D W A V E F O R M E L E C T R O N I C S S I N C E 1 9 7 5 info avtechpulse com http www avtechpuls...

Страница 2: ...ace said defective item This warranty does not apply to units which have been dissembled modified or subjected to conditions exceeding the applicable specifications or ratings This warranty is the ext...

Страница 3: ...ATION 10 VISUAL CHECK 10 POWER RATINGS 10 CONNECTION TO THE POWER SUPPLY 10 PROTECTION FROM ELECTRIC SHOCK 11 ENVIRONMENTAL CONDITIONS 12 LABVIEW DRIVERS 12 FUSES 13 AC FUSE REPLACEMENT 13 DC FUSE REP...

Страница 4: ...INTERFERENCE 31 MAINTENANCE 32 REGULAR MAINTENANCE 32 CLEANING 32 TRIGGER DAMAGE 32 WIRING DIAGRAMS 33 WIRING OF AC POWER 33 PCB 158R5 LOW VOLTAGE POWER SUPPLY 34 PCB 235C HIGH VOLTAGE DC POWER SUPPL...

Страница 5: ...0 load The AVR E3A B is a highly flexible instrument Aside from the internal trigger source it can also be triggered or gated by external TTL level signals A front panel pushbutton or a computer comma...

Страница 6: ...pically 3 of setting 5 ns Pulse width Typically 3 of setting 2 ns at maximum amplitude As the amplitude is reduced the pulse width may shift 5 ns For high accuracy applications requiring traceable cal...

Страница 7: ...g Also check that your load is adequately shielded It may be necessary to enclose the load in a metal enclosure If any of the connectors on the instrument are unused they should be covered with shield...

Страница 8: ...tion of conformity is issued under the sole responsibility of Avtech Electrosystems Ltd Specifically products manufactured do not contain the substances listed in the table below in concentrations gre...

Страница 9: ...9 have been made Requirements for WEEE compliance may include registration of products with local governments reporting of recycling activities to local governments and financing of recycling activiti...

Страница 10: ...mal single phase supply CONNECTION TO THE POWER SUPPLY An IEC 320 three pronged recessed male socket is provided on the back panel for AC power connection to the instrument One end of the detachable p...

Страница 11: ...erators are prevented access and or are insulated from every connection point In some cases connections must be exposed to potential human contact Operators must be trained to protect themselves from...

Страница 12: ...C decreasing linearly to 50 relative humidity at 40 C 5 Mains supply voltage fluctuations up to 10 of the nominal voltage 6 no pollution or only dry non conductive pollution LABVIEW DRIVERS A LabVIEW...

Страница 13: ...the fuses on the rear panel are shown in the figure below AC FUSE REPLACEMENT To physically access the AC fuses the power cord must be detached from the rear panel of the instrument The fuse drawer ma...

Страница 14: ...Nominal Mains Voltage Rating Case Size Recommended Replacement Part Littelfuse Part Number Digi Key Stock Number 1 2 AC 100 240V 0 5A 250V Time Delay 5 20 mm 0218 500HXP F2416 ND 3 DC N A 1 6A 250V Ti...

Страница 15: ...ed for about 5 seconds after which the instrument will attempt to re enable the output i e light green for about 1 second If the overload condition persists the output will be disabled again i e light...

Страница 16: ...is decreases the setting by a factor of ten If one of the adjustable numeric parameters is displayed and this parameter can be both positive or negative this changes the sign of the parameter EXTRA FI...

Страница 17: ...f the instrument This input can be either active high or active low depending on the front panel settings or programming commands The instrument triggers normally when this input is unconnected When s...

Страница 18: ...ntrol internally See the Programming Manual for B Instruments for more details on RS 232 control 8 Network Connector This Ethernet connector allows the instrument to be remotely controlled using the V...

Страница 19: ...set to a positive value the SYNC pulse precedes the OUT pulse When the delay is set to a negative value the SYNC pulse follows the OUT pulse These pulses are illustrated below assuming internal trigg...

Страница 20: ...ment controls the trigger frequency and generates the clock internally External Trigger the instrument is triggered by an external TTL level clock on the back panel TRIG connector Manual Trigger the i...

Страница 21: ...nel gate menu or the appropriate programming commands This input can also be set to act synchronously or asynchronously When set to asynchronous mode the GATE will disable the output immediately Outpu...

Страница 22: ...pplied to this connector which is connected to the output centre conductor through a high quality RF inductor Do not exceed 50V 250 mA DC OFFSET OT OPTION This option adds a bias tee circuit on the ou...

Страница 23: ...ned off when the instrument is not in use In the case of failure the switching elements are easily replaced following the procedure described in a following section DO NOT EXCEED 200 kHz The output st...

Страница 24: ...roper operation The output stage may be damaged if the output is not terminated into a 50 load The AVR E3A B output may be observed using either a high speed sampling oscilloscope or a slower real tim...

Страница 25: ...some sampling oscilloscopes a 20 dB attenuator should be placed at the input to the scope trigger channel d The bandwidth capability of components and instruments used to display the pulse generator o...

Страница 26: ...ow pointer should be pointing at the Normal choice If it is not press MOVE until it is d Press CHANGE to return to the main menu 5 To set the pulse width to 100 ns a Press the MOVE button until the ar...

Страница 27: ...ting to a value more compatible with your sampling oscilloscope Repeat step 4 if required You may also need to adjust the sampling scope controls d Rotate the ADJUST knob The amplitude as seen on the...

Страница 28: ...turns on the output For triggering a single event this sequence would be more appropriate rst resets the instrument trigger source hold turns off all triggering pulse width 50 ns sets the pulse width...

Страница 29: ...ed numeric value SOURce PULSe PERiod numeric value WIDTh numeric value DCYCle numeric value HOLD WIDTh DCYCle DELay numeric value GATE TYPE ASYNC SYNC LEVel HIgh LOw SOURce VOLTage LEVel IMMediate AMP...

Страница 30: ...nly VERSion query only TRIGger SOURce INTernal EXTernal MANual HOLD IMMediate CLS no query form ESE numeric value ESR query only IDN query only OPC SAV 0 1 2 3 no query form RCL 0 1 2 3 no query form...

Страница 31: ...factory for repair Service is to be performed solely by qualified service personnel Caution High voltages are present inside the instrument during normal operation Do not operate the instrument with t...

Страница 32: ...s recommended TRIGGER DAMAGE The rear panel TRIG input used in the external trigger mode is protected by a diode clamping circuit However the protection circuit is not foolproof and it is possible for...

Страница 33: ...WIRING DIAGRAMS WIRING OF AC POWER...

Страница 34: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...

Страница 35: ...PCB 235C HIGH VOLTAGE DC POWER SUPPLY...

Страница 36: ...PCB 231A DC OFFSET SOURCE...

Страница 37: ...PCB 217B RELAY DRIVER...

Страница 38: ...PCB 104G KEYPAD DISPLAY BOARD...

Страница 39: ...1 OFFSET 0 10V 2 PW 0 10V 3 SPARE 0 10V 4 O POL 5 O ZOUT PWRNG 6 O LOGIC 7 O SQU 8 O TRI 9 O SINE 10 O EA 11 O A LIVE 12 AMPLRNG4 13 AMPLRNG3 14 AMPLRNG2 15 AMPLRNG1 16 AMPLRNG0 17 XRLY 5 18 XRLY 4 E...

Страница 40: ...PERFORMANCE CHECKSHEET 40...

Отзывы: