FRONT PANEL CONTROLS
1. POWER Switch
. This is the main power switch. When turning the instrument on,
there is normally a delay of 10 seconds before anything is shown on the main
display, as the internal operating system boots up.
If the main menu does not appear after 30 seconds, turn off the instrument and
leave it off for at least 60 seconds before applying power again.
2. OVERLOAD Indicator
. When the instrument is powered, this indicator is normally
green, indicating normal operation. If this indicator is yellow, an internal automatic
overload protection circuit has been tripped. If the unit is overloaded (by operating
at an exceedingly high duty cycle or by operating into a very low impedance), the
protective circuit will disable the output of the instrument and turn the indicator light
yellow. The light will stay yellow (i.e. output disabled) for about 5 seconds after
which the instrument will attempt to re-enable the output (i.e. light green) for about
1 second. If the overload condition persists, the output will be disabled again (i.e.
light yellow) for another 5 seconds. If the overload condition has been removed, the
instrument will resume normal operation.
This overload indicator may flash yellow briefly at start-up. This is not a cause for
concern.
3. SYNC OUT
. This connector supplies a SYNC output that can be used to trigger
other equipment, particularly oscilloscopes. This signal leads (or lags) the main
output by a duration set by the "DELAY" controls and has an approximate
amplitude of +3 Volts to R
L
> 1kΩ with a pulse width of approximately 100 ns.
4. LIQUID CRYSTAL DISPLAY (LCD)
. This LCD is used in conjunction with the
keypad to change the instrument settings. Normally, the main menu is displayed,
which lists the key adjustable parameters and their current values. The
“Programming Manual for -B Instruments” describes the menus and submenus in
detail.
3
4
5
16
2
1
Содержание AVO-6HF-B
Страница 37: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...
Страница 39: ...PCB 197G HIGH VOLTAGE DISCHARGE BOARD...
Страница 48: ...PERFORMANCE CHECK SHEET 48...