17
·
Always wash the baffle after each use.
·
Do not use a brush or pin etc., to clean the parts.
·
When disinfecting the parts by boiling, be sure to boil them in plenty
of water.
·
Do not boil the baffle together with other objects, other than
applicable nebulizer accessories.
·
Do not use a microwave oven, dish dryer or hair dryer to dry parts.
CAUTION
Daily Disinfecting
Disinfect the nebulizer kit and particle adjusters and mask after the last
treatment of the day.
To disinfect the parts, use one of the methods described below:
A. Use a commercially available disinfectant. Follow the instructions provided
by the disinfectant manufacturer.
1. Submerge the parts in the cleansing solution for the specified period.
2. Remove the parts and discard the solution.
3. Rinse the parts with clean hot tap water, shake off excess water and allow
to air dry in a clean environment.
B. Parts may be boiled between 15 to 20 minutes. After boiling, carefully
remove the parts, shake off excess water and allow to air dry in a clean
environment.
NOTE: Do not boil air tube, air filter, air filter cover and optional masks.
Do not use an autoclave, EOG gas disinfection or low temperature
plasma sterilizer to disinfect the device.
CAUTION
Handling the Baffle
The baffle is an important part that is used to nebulize medication.
Be sure to observe the following precautions when handling it.