Model 6700 Family User’s Manual
20
5.
When the desired capacity is displayed,
press the
TEST
key. . .
That choice is accepted and
LOAD 0
is
displayed.
6.
Clear all weight from the scale platter and
press the
TEST
key…
After a brief wait
LOAD xx
is displayed.
Alternate calibration points can be
chosen using the
ZERO
key to scroll
between choices (see Table 1).
7.
Place the appropriate calibration weights on
the scale and press the
TEST
key. After a
brief wait…
DONE
is displayed.
8.
Remove all calibration weights from scale.
9.
Press the
TEST
key…
DIAG
is displayed, or return Switch 1 to
the closed position. The scale returns to
normal weighing mode.
The scale is now tested, configured, and cali-
brated. It is ready for use in your application.
The capacity selected
must correlate with the
rated capacity of the
scale noted on the
serial tag.
If this procedure is
attempted without any
calibration weights
applied, the scale will
abort the process and
retain the original
calibration data.