Chapter 3: System Operation
KXL System Operator’s Manual,
Rev G
3-5
Figure 3-3. Treatment Parameters Screen
Specify the desired treatment parameters, in turn, for the patient as follows:
o
UV Treatment Time (1 – 30 minutes)
o
UV Power (3 – 45 mW/cm
2
)
NOTE: Total Energy will be displayed in the lower orange box. The maximum
permissible total Energy density is 10 J/cm
2
.
When finished entering treatment parameters, press the
Checkmark
button.
Figure 3-4. Confirm Treatment Parameters Screen
Confirm the specified treatment parameters by pressing the Checkmark; if the treatment
parameters are not correct press the X and then re-enter the desired treatment parameters.