24-08-04
AvaLight-HAL manual.doc
1
Avantes
www.avantes.com
AvaLight-HAL Tungsten Halogen light source Operating Manual
PS-24VDC power supply
Starting up
1.
Plug in the PS-24V power supply .
2.
Plug in the connector of the power supply into the socket of AvaLight-HAL.
3.
Connect the SMA-connectors of your fibers to the SMA-plug
4.
Switch on the Halogen lamp with the switch .
Settings for high power or long lifetime
The AvaLight-HAL is factory set for optimal power output (medium setting), with an internal jumper the optical
energy can be controlled (see table).
Jumper setting
Optical output
Color temperature
Exp. Lifetime
LOW
70%
Ca. 2.700K
> 2.000 hrs.
MEDIUM (default)
100%
Ca. 2.850K
1.000 hrs
HIGH
150%
Ca. 3.000K
< 1.000 hrs.
The internal jumper settings can be changed, please follow instructions:
1.
Remove screw protection caps on the backside
2.
Loosen 2 screws with philips screwdriver
3.
remove backplane
4.
take out electronics board
5.
Adjust jumper settings according to scheme below
6.
put back electronics board, be carefull not to pinch the
electrical wires
7.
put back backplane, screws and screw protection caps
SMA plug
24V Power supply
On-Off switch
Screw caps
Jumper JP1