2

 

AvaLight-D(H)-S  

2.1

 

Quick start 

 

 

 

 

SMA-plug 

Connect the SMA connector of the fiber to the SMA-plug of the Light Source. 

 

Connect your Fiber optic Cable first before starting the Deuterium Lamp. Please be sure to 

avoid direct radiation of skin and eyes at the fiber output.  

 

 

 

TTL shutter

 

Shutter LED

 

SMA-plug

 

Halogen switch

 

Halogen LED

 

Deuterium 
switch

 

Deuterium LED

 

Power 
connector

 

Double fuse

 

Power switch

 

I/O connector

 

Halogen 
adjustment port

 

Содержание AvaLight-D-S

Страница 1: ......

Страница 2: ...oduct specifications and descriptions in this manual are subject to change without notice Avantes BV will not be responsible for errors and omissions in this manual or for direct or incidental damages...

Страница 3: ...Shutter Open 5 2 5 Power connector with double Fuse and switch 5 2 6 Halogen light power adjustment 6 2 7 I O connector 6 3 Replacement of Deuterium or Halogen Lamp 7 4 Technical support 9 4 1 Getting...

Страница 4: ...DUV version starts even lower at 175 nm for your deep UV experiments This version also offers twice the intensity at 200 nm The output of the AvaLight DH S is optimized for fibers or bundles upto 600...

Страница 5: ...the SMA plug of the Light Source Connect your Fiber optic Cable first before starting the Deuterium Lamp Please be sure to avoid direct radiation of skin and eyes at the fiber output TTL shutter Shut...

Страница 6: ...tching the Halogen switch downwards to ON the Halogen lamp switches on The Halogen LED indicates the operation mode Switching the Halogen switch upwards to OFF the Halogen lamp switches off Switch pos...

Страница 7: ...he halogen power 2 7 I O connector For automatic shutter control to be controlled by software it is needed to connect an AvaSpec spectrometer to the AvaLight D H S with the interface cable IC DB26 2 N...

Страница 8: ...he lamp envelope quartz glass should not be touched with bare fingers otherwise lifetime will be severely reduced Please follow these instructions for the bulb replacement 1 Switch off the light sourc...

Страница 9: ...crews to secure the deuterium lamp 12 Remove the 2 screws of the halogen lamp fitting 13 Take out the halogen lamp fitting 14 Replace the halogen bulb with new 5W bulb caution do not touch the lamp gl...

Страница 10: ...ches are ON before main power is switched on Put buttons for Deuterium and Halogen in OFF position before switching on main power 4 3 2 Deuterium Lamp Only FAULT POSSIBLE CAUSE REMEDY Power switch on...

Страница 11: ...gen Lamp Only FAULT POSSIBLE CAUSE REMEDY Halogen Lamp does not work after switching on Halogen Lamp defective Replace Halogen Lamp LED does not light up after switching on Halogen lamp Internal Power...

Страница 12: ...0 mm ca 5 kg total power for the specified wavelength range 5 1 Ordering Information AvaLight D S Deuterium light source 190 400 nm incl TTL shutter SR fibers needed AvaLight DH S Deuterium Halogen li...