Instructions and user manual ChloriProtect
NOT004ENV05
9
1)
Remove the protective
cap of the probe and keep it
for reuse at the time of
wintering.
2)
During 20 minutes, let soak the
probe to rinse it. We suggest you
to start with this step.
3)
Immerse the electrode in
the 475mV solution, stir and
leave it in the solution
without
touching
the
electrode or the cable
4)
After waiting minimum one minute, the stabilized value is
displayed. Press
CAL
until the display shows CAL flashing then
release. Then the quality of the probe is briefly displayed.
7.0 AFTER SALES SERVICE
To contact our Technical Services (contact details on page 1), you will need the following informations:
Type of device
Serial number
Device code
For any return of the unit, thank you to please follow the instructions in "
1.2 SHIPPING AND TRANSPORTING
"
on page 1 of this manual.