Avacen HOME XL Скачать руководство пользователя страница 7

20220913 

AVACEN HOME XL, AVACEN R3 HOME XL User Guide 

7.

 

IMPORTANT SAFETY INFORMATION 

 

Read the entire AVACEN device User Guide before using this device. 

 

For continued protection against risk of fire and shock, use the hand unit only with the 
provided power supply, AVACEN part number 50-100000. 

 

Do not store the AVACEN device

 

in

 

direct sunlight.

 

The AVACEN should be stored in a clean, 

well-ventilated area at room temperature. Keep out of reach of small children and pets.  
Optional carrying bags should be left open to allow for ventilation. 

 

Do not use outdoors in the rain, snow, or under water. While the device is protected against 
vertically dripping water, it is not water resistant, and should not be used in wet areas. 

 

Pressing too hard on the heat pad can cause the AVACEN to be ineffective. 

 

The Power Supply Connector is the main power disconnect device for the AVACEN device. If a 
malfunction is suspected, unplug the power supply or remove the power input from the unit. 

 

This device complies with Part 18 of the FCC Rules, as well as international electromagnetic 
compatibility requirements. Use is subject to these two conditions:.  
1) This device may not cause harmful interference, and 2) This device must accept any 
interference received, including interference that may cause undesired operation.  
These limits are designed to provide reasonable protection against harmful interference in a 
residential installation. However, there is no guarantee that interference will not occur in a 
particular installation. If this equipment causes problems with radio or TV reception, which 
can be checked by turning the equipment off and on, you are encouraged to try to correct the 
interference by one or more of the following measures:. 

• Reorient the receiving antenna 

 

• Relocate the system with respect to the receiver 

 

• Move the system away from the receiver

, and/or  

• Plug the system into a different outlet so that the syste

m and the receiver are on different 

branch circuits. 
 

 

 
 

 

 

 

 

Environmental Conditions 

Temperature 

Operating  50

F (10°C) to 104

(40°C) 

Storage  32

F (0°C) to 140

F (60°C) 

Relative 
Humidity 

Operating  30% to 75% rH 

Storage  10% to 95% rH 

Atmospheric 
Pressure 

Operating  700hPa to 1060hPa 

Storage  500hPa to 1060hPa 

 
AC Input Voltage 

HOME XL - 240V~, 50-
60Hz 

Expected service life of equipment 
and accessories 

5 years 

Содержание HOME XL

Страница 1: ...Registered Trademark of AVACEN Medical Inc Protected by one or more of the following U S patents 8 679 170 9 066 781 9 192 509 9 687 385 10 537 464 and international equivalents Other patents pending...

Страница 2: ...nnector Hand Positioning Tips KEEP WRIST STRAIGHT AND IN THE CENTER OF THE CUFF POSITION WRIST IN MIDDLE OF CUFF PARALLEL WITH CHAMBER CORRECT CORRECT INCORRECT INCORRECT Sanitary Mitt Metal Heat Pad...

Страница 3: ...ould treatment last 9 B How important is the placement of my hand 9 C Is it necessary to wear the AVACEN Mitt 9 D Do I need a new a Mitt every time I have a treatment 9 E Do I need to purchase the Mit...

Страница 4: ...13 U Sweating when using the AVACEN 13 V Discomfort in wrist or forearm during use 13 20 SETTINGS 14 A Language Selection 14 B Game Mode On Off 14 C Change Defaults 14 D Usage Statistics 15 E Hardware...

Страница 5: ...oam in case the AVACEN needs to be returned Be sure to read the What to Expect flyer before the first use 3 NORMAL USE The AVACEN is designed to provide dry heat therapy Treatment time and or number o...

Страница 6: ...harmful effect occurs while using the product stop use and contact your health care professional Use with care May cause serious burns Do not use on sensitive skin areas or in the presence of poor ci...

Страница 7: ...l electromagnetic compatibility requirements Use is subject to these two conditions 1 This device may not cause harmful interference and 2 This device must accept any interference received including i...

Страница 8: ...ow Medium High Temperature Settings Remove all jewelry and watch before inserting hand in the device Do not allow your shirt sleeve to enter into the Inflatable Cuff Caution Refer to Instructions Read...

Страница 9: ...ive at the Please Wait Warming up screen This message will remain until the hand pad has reached 98 6 F When the hand pad reaches the predetermined standby temperature the home screen will be displaye...

Страница 10: ...e off by mistake or on purpose remove your hand from the Vacuum Chamber before powering the device back on again 11 CUFF SEALING The Pressurizing Calibrating message is displayed while the vacuum moto...

Страница 11: ...gain 16 END OF THERAPY When the treatment time runs down to zero the device will finalize the session The cuff will begin to deflate and the Remove Hand Slowly message is displayed Remove your hand fr...

Страница 12: ...allied during the game positioned below the HPI The number of points earned is based on how well the user keeps the HPI a solid color Points are scored as long as the Lift Up message or the Press down...

Страница 13: ...only Any level of major discomfort after adjusting temperature is not normal Stop using the AVACEN right away and report the problem to AVACEN at 888 4 AVACEN 888 428 2236 and select Customer Service...

Страница 14: ...rly positioned See section 22 PERIODIC MAINTENANCE for taking the Inflatable Cuff on and off If your skin is very dry put a small amount of hand lotion on your wrist and rub it in to improve sealing a...

Страница 15: ...ase the onboard computer monitors if there is too much or too little palm pressure against the heat pad A warning will appear on the display if too little or too much hand pressure has been applied Af...

Страница 16: ...d will cause the Lift Hand Slightly message to appear along with a beeping sound At this point just retract your hand into the cuff to its original position and the alerts will stop To make it easier...

Страница 17: ...R Red or purple spots on the hand Red or purple spots on the hand usually the backside after the first couple of AVACEN device uses can be caused by a minor hemorrhage petechia These spots are harmle...

Страница 18: ...ent selection is underlined Choose the desired language and it will become underlined Press Go Back to return to settings menu B Game Mode On Off From the Settings Menu tap the Game Mode On Off line t...

Страница 19: ...alibration locked iii Force Sensor Calibration locked iv Force Measurement Test v Vacuum Level Average F Last Failure This menu stores information about errors when the device was last used If you are...

Страница 20: ...setting aside for reassembly Next lift up four tabs of the lock ring circled in the images below two on top two on bottom A tool has been provided to access some of the more difficult snaps A video i...

Страница 21: ...om the tabs as the old cuff This may require some levering of the gasket away from the cuff to fully overlap on the product side Gently slip lock ring over the cuff Ensure four snaps are heard one for...

Страница 22: ......

Страница 23: ...Pressure Holes Power Supply Connector Hand Positioning Tips KEEP WRIST STRAIGHT AND IN THE CENTER OF THE CUFF POSITION WRIST IN MIDDLE OF CUFF PARALLEL WITH CHAMBER CORRECT CORRECT INCORRECT INCORRECT...

Страница 24: ...rom Home screen Notes a Jewelry must be removed on treatment hand wrist b Default Time Temp is 20 Min High c If Temp is too hot select a lower temp and work back to High over time Warnings Contact a p...

Отзывы: