Versión 08 Enero 2018
Pág. - 36 -
in the left bottom of the right side of the machine after connection. The red light on the
control panel will be on after a short buzz of the buzzer. It will display the main interface
after self-examination. Then, proceed with next step.
4.3 Open the door
Press the
“
DOOR/STOP
“
button of the control panel and the door cover will eject
automatically to a certain height. Then, you need to lift the cover to open the door
completely. You will see the cavity of the centrifuge. You can remove the rotor and
package for the first use of the new machine.
Note: the door cover will spring to certain height automatically while opening,
please prevent your head or anything else from the upward side of the door cover to
avoid dangers! You can open the door manually in case it goes wrong and the door
cover can't open automatically, but you have to remove the materials from the centrifuge.
Please check
“
troubleshootin
g”
for more details.
4.4 Close the door
Press the door cover until the door hook in its upper front hooks the lock dowel and
you hear a
“
click
”
. The bottom of the door hook will touch the over travel-limit switch and
then the door will be locked.
Please press the door cover properly. Don't overexert,
or the hook will be damaged.
4.5 Install the rotor
The rotor should be confirmed and designated by Nahita, otherwise it may cause
problems. The rotor designated by Nahita for this centrifuge is the Angle rotor 12x15 mL.
Users are advised to use the designated rotor in this manual.
Inappropriate types of rotors and centrifuge tubes
will lead to a poor effect, or even the damage of the
centrifuge.