33
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Surge
± 1 kV
line(s) to
line(s)
±1 kV
differential
Main power
quality should be
that of a typical
IEC 61000-
4-5
Mode
Commercial or
hospital
environment.
Power
frequency
(50Hz/60Hz)
magnetic
field
IEC 61000-
4-8
30A/m
Power frequency
magnetic
fields should
be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
NOTE UT is the ac. mains voltage prior to application of
the test level.