APT3100
Smart Pressure Transmitter
Operation Manual M3100-EO1G
23
Duon System Co.,Ltd.
3.8 Shop Commissioning using HHT
Commissioning consists of testing the transmitter, testing the loop, and verifying transmitter
configuration data. APT-3100 Pressure Transmitter may be commissioned using HHT of HART supported
either before of after installation.
▲
If you connect "TEST" pin, it's not communicated. If it doesn't exposed electronics circuits after
install, you must connect all Jumper of transmitter in the shop commissioning level. Analog output of
transmitter is 4~20 mA, thus it requires Power Supply at 11.9 V ~ 45 Vdc and Ampere meter for
measuring output current. You must connect 250~550 ohm resistor in Power Loop for HART
Communication and HHT or PC Configurator.
[Figure 3-5 Connection the transmitter to HHT]
www.