Communication Interface

  

■ RS485

Comm. protocol

Modbus RTU

Application standard

Compliance with EIA RS485 

Max. connection

31 units (address: 01 to 99)

Comm. synchronous method

Asynchronous

Comm. method

2-wire half duplex 

Comm. distance

≤ 800 m

Comm. speed

1,200 / 1,400 / 4,800 / 9,600 / 19,200 / 38,400 bps

Start bit

1-bit (fixed) 

Data bit

8-bit (fixed) 

Parity bit

None, Even, Odd

Stop bit

1-bit, 2-bit

EEPROM life cycle

≈ 1,000,000 operations (Erase / Write)

Specifications

Model

MT4Y-DV-

4□

MT4Y-DA-

4□

MT4Y-AV-

4□

MT4Y-AA-

4□

Input type

DC voltage

DC current

AC voltage 

01)

AC current 

 01)

Max.

 allowable 

input 

110 % F.S. for each measured input range

Display method

7-segment (red) LED (

c

haracter height: 14.2 mm)

Display accuracy

Dependent on the ambient temperature

23 ± 5 ℃

± 0.1 % F.S. rdg 

± 2 digit

± 0.1 % F.S. rdg 

± 2 digit 

 02)

± 0.3 % F.S. rdg 

± 3 digit

± 0.3 % F.S. rdg 

± 3 digit

-10 to 50 ℃

± 0.5 % F.S. rdg ± 3 digit

Max. display range

-1999 to 9999 (4 digit) 

A / D 

conversion 

method

ΣΔ (Sigma Delta) ADC

Sampling cycle

50 ms

16.6 ms

Unit weight 

(packaged)

≈ 134 g (≈ 213.5 g)

Approval

ᜢ ᜧ ᜫ

 01) Available frequency display, Display accuracy (23 ± 5 ℃): ± 0.1 % F.S. rdg ± 2 digit
 02)  5 A terminal: ± 0.3 % F.S. rdg ± 3 digit 

P

reset output

None (indicator) / Relay / NPN open collector / PNP open collector 

output model

Relay

Contact capacity: 250 VACᜠ 3 A, 30 VDCᜡ 3 A

Contact composition: N.O (1a)

NPN / PNP 

 

open collector

Output capacity: ≤ 12 - 24 VDCᜡ ± 2 VDCᜡ, 50 mA resistive load

Sub

 output

None (indicator) / BCD Dynamic / Transmission (DC 4 - 20 mA) / 

Low speed serial / RS485 Communication output model 

BCD Dynamic / 

 

Low speed

 serial 

NPN open collector output

Output capacity: ≤ 12 - 24 VDCᜡ, 50 mA resistive load

Transmission

(DC 4 - 20 mA)

Resolution: 1/12,000 (load resistance: ≤ 600 Ω)

Response time: ≤ 450 ms

RS485 communication Protocol: Modbus RTU

Power supply

100 - 240 VACᜠ ± 10 %  50 / 60 Hz

Power consumption

5 VA

Insulation resistance

B

etween external terminal and case: ≥ 100 MΩ (500 VDCᜡ megger)

Dielectric strength

B

etween external terminal and case: 2,000 VACᜠ 50 / 60 Hz for 1 min

Noise immunity 

± 2 kV square wave noise (pulse width: 1 ㎲) by the noise simulator

Vibration

0.75 mm double amplitude at frequency of 10 to 55 Hz (for 1 min) in 

each X, Y, Z direction for 2 hours

Vibration 

(malfunction)

0.5 mm double amplitude at frequency of 10 to 55 Hz (for 1 min) in 

each X, Y, Z direction for 10 min

Shock 

300 m/s

(≈ 30 G) in each X, Y, Z direction for 3 times

Shock (malfunction)

100 m/s

(≈ 10 G) in each X, Y, Z direction for 3 times

Relay life cycle

Mechanical

: ≥ 20,000,000 operations

Electrical: ≥ 100,000 operations (250 VACᜠ 3A resistive load) 

Ambient temp.

-10 to 50 ℃, storage: -20 to 60 ℃ (no freezing or condensation)

Ambient humi.

35 to 85 %RH, storage: 35 to 85 %RH (no freezing or condensation)

Insulation type

Symbol:  , double or reinforced insulation (dielectric strength between 

the measurement input part and the power part: 1 kV) 

Parameter Setting

• Some parameters are activated / deactivated depending on the model or setting of 

other parameters. Refer to the description of each parameter.

• If any key is not entered for 60 sec in each parameter, it returns to RUN mode.

• After returning to RUN mode, press the [MODE] key within 2 sec, it returns to 

previous parameter.

• [MODE] key: Saves current setting value and moves to the next parameter.

 

[◀] key: Checks fixed value / Changes setting digits.

 

[▲], [▼] key: Changes setting values.

  

■ Parameter 1 group

Parameter

Mark

Defaults

Setting range

Display 

condition

1-1 Input range

IN-R

500V

[DC voltage model], [AC voltage model]

• 

Refer to Input Range and Display Range

-

5A

[DC current model], [AC current model]

• 

Refer to Input Range and Display Range

-

1-2

Display 

method

DISP

STND

STND: standard, SCAL: 

scale

, FREQ: 

frequency

 01)

-

1-3

Measurement 

method

IN-T

tRMS

[AC voltage model], [AC current model]

T.RMS: True RMS, A.RMS: Average RMS, 

AVG

•  True RMS =

A

1

2

 + A

2

2

 + ··· + A

n

2

            n              

• 

A

verage RMS =  

A

1

 + A

2

 + ··· + A

n

            n            

 

× Waveform rate

(n = number of display values   per cycle, 

A = display value)

1-2 

Display 

method:

 

STND, 

SCAL

1-4

Max. display 

value 

(fixed)

STND

50)0

[DC voltage model], [AC voltage model]

Max. value of display range

1-2 

Display 

method: 

STND

%000

[DC current model], [AC current model]

Max. value of display range

1-5

High-limit 

display value 

gradient 

correction

INbH

!000

0.100 to 5.000 %

1-6

Low-limit 

display value 

deviation 

correction

INbL

00

-99 to 99

1-7

Decimal 

point 

position

DOT

)0

[DC voltage model], [AC voltage model]

0, 0.0, 0.00, 0.000

1-2 

Display 

method:

 

SCAL

&

* 1-7 

Decimal 

point 

position:

0.0, 0.00, 

0.000

)000

[DC current model], [AC current model]

0, 0.0, 0.00, 0.000

1-8 High-limit scale

H-SC

-

Display value against max. measurement input*

1-9 Low-limit scale

L-SC

-

Display value against min. measurement input*

1-10

High-limit 

display value 

gradient 

correction

INbH

!000

0.100 to 5.000 %

1-11

Low-limit 

display value 

deviation 

correction

 02)

INbL

00

w-99 to 99

1-12

Decimal 

point 

position

 03)

DOT

)0

[AC voltage model]

0, 0.0, 0.00, 0.000

1-2 

Display 

method:

 

FREQ

)000

[AC current model]

0, 0.0, 0.00, 0.000

1-13

High-limit 

display value 

gradient 

correction

INbH

!000

0.100 to 9.999

1-14

Exponent of 

INB

INbE

10-0

10-0: 10

0

, 10-1: 10

-1

, 10-2: 10

-2

, 10 1: 10

1

 01)  Displays at AC voltage or AC current model only.
 02) 

 

Low-limit display value deviation correction range is within -99 to 99 for D

0

, D

-1

 digit regardless of decimal point position.

 03)   Display range is variable according to decimal point position.

 

Dot

Display range

Frequency 

measurement range

0

-1999 to 9999

1 to 9999 Hz

)0

-199.9 to 999.9

0.1 to 999.9 Hz

)00

-19.99 to 99.99

0.10 to 99.99 Hz

)000

-1.999 to 9.999

0.100 to 9.999 Hz

Mode Setting

RUN

[MODE] 3 

sec

Parameter 1 

group

[MODE] 3 sec

RUN

[MODE] 5 

sec

Parameter 2 

group

[MODE] 3 sec

[MODE]

Parameter 0  

group

[MODE] 3 sec

[◀] + [▲] 3 sec

Zero adjustment

Auto

[◀] + [▲] + [▼] 5 sec

Reset

[MODE]

Отзывы: