2.3 4.75V

This section describes line and load characteristics of the 4.75V output which may be used to supply

external devices. As will be shown, up to 50 mA may be drawn from this output. This seems a reasonable

limit when keeping in mind that battery life with an external device should still amount to several hours.

The 4.75V output voltage is generated inside the 6150AD out of the battery voltage by means of a

switched regulator. It is individually tested in serial production to be in the range of 4.65V to 4.90V with a

typical value of 4.75V (at an input battery voltage of 7.0V).

no load

RL = 82.5 Ohm

U

in

V

U

out

V

I

in

mA

U

out

V

I

out

mA

efficiency

%

5.0

4.782

52.0

4.320

52.4

87

5.5

4.785

55.0

4.651

56.4

87

6.0

4.789

52.5

4.683

56.8

84

7.0

4.795

46.5

4.724

57.3

83

8.0

4.801

42.0

4.744

57.5

81

9.0

4.807

38.0

4.758

57.7

80

9.5

4.810

37.0

4.765

57.8

78

Note that at input battery voltages below 5.5 volts the 6150AD will issue battery warning.

End of »Data Outputs« section.

Appendix with »Data Inputs« section is only supplied on special request.

in

I

I out

U in

U out

RL

6150 AD

4.75V

GND

06/2003

Technical Manual 6150AD Probe Connector

Page 7 of 14

Daimlerstrasse 27

Telefon  +49(0)6203-9503-00 http://www.automess.de

USt.-ID-Nr. / VAT ID No.

D-68526 Ladenburg Telefax   +49(0)6203-9503-29 eMail: [email protected] DE 144 458 420

Содержание 6150AD

Страница 1: ...Data Inputs 8 3 1 PL and S E Lines 8 3 2 Ext Input 10 3 3 Probe Characteristics 11 Daimlerstrasse 27 Telefon 49 0 6203 9503 00 http www automess de USt ID Nr VAT ID No D 68526 Ladenburg Telefax 49 0...

Страница 2: ...low an will be discussed in more detail later GND Ground potential Note that there is no pin for GND The cable shield serves both as shield and as ground connection The aluminium housing of the 6150AD...

Страница 3: ...ed Ohms Its driving capability is therefore at least equivalent to RS232 interfaces Nevertheless cables should be restricted to maximum lengths as with RS232 interfaces 15m at 9600Bd 30m at 4800Bd etc...

Страница 4: ...xponent 6 block check character The second byte type of detector 6150AD contains Bit 0 5 detector in use 0 probe 6150AD 0 7 probe 6150AD b 15 probe 6150AD 15 17 probe 6150AD 17 18 probe 6150AD 18 19 p...

Страница 5: ...e of detector 6150AD bc sonde mantlo ASC INPUT 1 1 low order mantissa bc bc XOR mantlo manthi ASC INPUT 1 1 high order mantissa bc bc XOR manthi expon ASC INPUT 1 1 exponent bc bc XOR expon IF expon 1...

Страница 6: ...igh indicates that dose rate alarm is off and Control low indicates that dose rate alarm is on This allows triggering of external alarming devices However one has to consider that the Control output c...

Страница 7: ...put battery voltage of 7 0V no load RL 82 5 Ohm Uin V Uout V Iin mA Uout V Iout mA efficiency 5 0 4 782 52 0 4 320 52 4 87 5 5 4 785 55 0 4 651 56 4 87 6 0 4 789 52 5 4 683 56 8 84 7 0 4 795 46 5 4 72...

Отзывы: