Automess 6150AD Скачать руководство пользователя страница 1

Contents

1. General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2

. . . . . . .

2. Data Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3

. . . . . . .

2.1 »Term« . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3

. . . . . . .

2.1.1 Hardware Characteristics . . . . . . . . . . . .

3

. . . . . . .

2.1.2 Software Characteristics . . . . . . . . . . . . .

4

. . . . . . .

2.2 »Control« . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6

. . . . . . .

2.3 4.75V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7

. . . . . . .

3. Appendix: Data Inputs . . . . . . . . . . . . . . . . . . . . . . . . .

8

. . . . . . .

3.1 »PL« and »S/E« Lines . . . . . . . . . . . . . . . . . . . . . .

8

. . . . . . .

3.2 »Ext« Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

. . . . . .

3.3 Probe Characteristics . . . . . . . . . . . . . . . . . . . . . .

11

. . . . . .

Daimlerstrasse 27

Telefon  +49(0)6203-9503-00 http://www.automess.de

USt.-ID-Nr. / VAT ID No.

D-68526 Ladenburg Telefax   +49(0)6203-9503-29 eMail: [email protected] DE 144 458 420

 

Automation und Messtechnik GmbH

23 June 2003

Dose Rate Meter 6150AD: Probe Connector

Technical Manual

Содержание 6150AD

Страница 1: ...Data Inputs 8 3 1 PL and S E Lines 8 3 2 Ext Input 10 3 3 Probe Characteristics 11 Daimlerstrasse 27 Telefon 49 0 6203 9503 00 http www automess de USt ID Nr VAT ID No D 68526 Ladenburg Telefax 49 0...

Страница 2: ...low an will be discussed in more detail later GND Ground potential Note that there is no pin for GND The cable shield serves both as shield and as ground connection The aluminium housing of the 6150AD...

Страница 3: ...ed Ohms Its driving capability is therefore at least equivalent to RS232 interfaces Nevertheless cables should be restricted to maximum lengths as with RS232 interfaces 15m at 9600Bd 30m at 4800Bd etc...

Страница 4: ...xponent 6 block check character The second byte type of detector 6150AD contains Bit 0 5 detector in use 0 probe 6150AD 0 7 probe 6150AD b 15 probe 6150AD 15 17 probe 6150AD 17 18 probe 6150AD 18 19 p...

Страница 5: ...e of detector 6150AD bc sonde mantlo ASC INPUT 1 1 low order mantissa bc bc XOR mantlo manthi ASC INPUT 1 1 high order mantissa bc bc XOR manthi expon ASC INPUT 1 1 exponent bc bc XOR expon IF expon 1...

Страница 6: ...igh indicates that dose rate alarm is off and Control low indicates that dose rate alarm is on This allows triggering of external alarming devices However one has to consider that the Control output c...

Страница 7: ...put battery voltage of 7 0V no load RL 82 5 Ohm Uin V Uout V Iin mA Uout V Iout mA efficiency 5 0 4 782 52 0 4 320 52 4 87 5 5 4 785 55 0 4 651 56 4 87 6 0 4 789 52 5 4 683 56 8 84 7 0 4 795 46 5 4 72...