background image

Instruction Manual QNix

®

8500 

- 2 - 

Содержание QNix 8500

Страница 1: ......

Страница 2: ...Instruction Manual QNix 8500 2...

Страница 3: ...Instruction Manual QNix 8500 3...

Страница 4: ...Averaging 11 8 3 Data Memory 11 8 3 1 Delete All 12 8 3 2 Delete values 12 8 3 3 Data Memory Mode ON OFF 12 8 3 4 Creating a New Batch 12 8 3 5 Selecting Editing a Batch 12 8 4 Statistics 13 8 4 1 Se...

Страница 5: ...e PC Software 17 14 Loading a new Language into the Gauge 17 15 Updating the Software in the Gauge and Probe 18 16 Display Symbols 18 17 Scope of Delivery 19 18 Technical Data 20 Notice This instructi...

Страница 6: ...to strong magnetic fields Remove the batteries from the gauge and store them separately if the device is not used for a longer period The gauge is not waterproof Do not expose it to water 3 Changing...

Страница 7: ...ring head Within the measuring range of the probe measurements like e g distance measurements are possible that way 6 Setting the Zero Reference When using the gauge for the first time after changing...

Страница 8: ...u For your convenience find a diagram of the menu structure showing all possible menu op tions on the first two pages of this instruction manual Depending on the version of your gauge only a part of t...

Страница 9: ...d calibrations under the menu item Define New Calibration To select a certain calibration use the and buttons Subse quently pressing opens the corresponding options menu A dot at the right side of the...

Страница 10: ...a number of zero reference readings max 19 to be taken with the probe before the gauge will calculate the average zero reference based on the readings taken After selecting the menu item pressing the...

Страница 11: ...t To set or delete the lower limit please follow the instructions given under 8 2 3 Readings below the lower limit are indicated in the display with the symbol and with an au dio warning signal 8 2 5...

Страница 12: ...firm it by pressing For corrections of prior digits simply press the backspace button at any time NOTE When creating a new batch after other batches i e B 0001 B 0002 have been de leted the gauge stil...

Страница 13: ...he statistics functions are calculated per memory block The gauge determines the following statistical parameters average value standard deviation maximum minimum Within the statistics menu two differ...

Страница 14: ...ing of the display To change the display orientation press the button then select the setting Normal or Flip using and and finally confirm by pressing again NOTE When the display is flipped the functi...

Страница 15: ...he and buttons Confirm with For corrections use the backspace button 8 5 6 2 Setting the Date Set the date the same way you set the time 8 5 7 The button The button is used to set the zero reference s...

Страница 16: ...coating you want to measure If an uncoated piece is unobtainable you can use the standard plates that are supplied with your gauge Refer to 6 for further details Please note that it is the probe that...

Страница 17: ...tion guide on the CD 13 Setting the Gauge Options with the PC Software The setup options described in 6can also be programmed and then downloaded into the gauge using the PC software This is especiall...

Страница 18: ...rror operating error Infi readings outside measuring range wrong type of surface readings exceeding upper limit readings below lower limit x average standard deviation wireless interface active p radi...

Страница 19: ...th clip Plastic case for transportation and storage Optional Accessories Fe probe 2 mm Fe probe 5 mm NFe probe 2 mm Dual probe Fe NFe 2 2 mm Dual probe Fe NFe 5 2 mm Calibration films zero reference p...

Страница 20: ...max 10 m on the free field Temperature range Storage 10 C to 60 C 14 F to 140 F Operation 0 C to 50 C 32 F to 122 Power supply 2 x batteries AA 1 5V alkaline or 2 x rechargeable batteries AA 1 2V Dime...

Страница 21: ...e pole and negative pole to avoid short circuiting the device Do not use new and old batteries or batteries of a different kind together This could lead to miss functions in your device Remove empty b...