Hold the gauge according to the drawing and slightly press the 
probe onto the surface so that the ring around the probe is 
evenly placed. If the value indicated on the zero plate or on 
your uncoated substrate is not within the basic tolerance, the 
gauge should be adjusted (gauges with a memory must not be 
in the Memory Mode). Push one of the slides at the sides of 
the gauge while the probe is placed firmly on the substrate, i.e. 
the zero plate. You will hear a ”Beep“ and a control number 
will appear on the display. Lift the gauge at least 25 mm or 
one inch from the zero plate. The gauge will ”beep“ again and 
another number will be displayed. The 

QNix® 1500 

is 

adjusted now.  
When repeating measurements on the same spot, the 
measured value may not always be 0.00, due to surface 
roughness, dirt, scratches etc. 
The gauge is delivered with two protective rings around the 
probe for better positioning and protection. These rings can 
easily be removed by slightly turning and pulling. On concave 
surfaces remove the protective rings in order to achieve 
accurate measurements. 

Measurement 

Place the 

QNix® 

1500

 with its ring 

around the probe 
plane on the 
object to be 
measured (see 
drawing). For 
measurements on 
steel or iron, the 
Fe-sticker should 
face you. For 
measurements on 
non-magnetic metallic substrates such as aluminum you 
should be able to see the NFe-sticker. 
Do not place your finger on the upper probe, otherwise the 
readings will not be correct. Take care that the circular probe 
ring is placed evenly on the surface. On even objects, the two 
rear supporting points will help to position the gauge. The 

5

Содержание QNix 1500

Страница 1: ...1 Instruction Manual Coating Thickness Gauge 1500 Optionally with Memory...

Страница 2: ...s this gauge offers Would you like information on the other gauges in our coating thickness measurement line or do you have any questions concerning coating thickness measurement Please do not hesitat...

Страница 3: ...adings as it is the case with all precision gauges Do not expose the gauge to strong direct sunlight or to other sources of energy causing temperature shocks Protect the gauge from moisture chemically...

Страница 4: ...ttery The battery is located at the wide end of the gauge If the blinking sign BAT appears on the display it is advisable to change the battery but it is still possible to measure for some time Memori...

Страница 5: ...on the same spot the measured value may not always be 0 00 due to surface roughness dirt scratches etc The gauge is delivered with two protective rings around the probe for better positioning and prot...

Страница 6: ...th a built in memory and an interface RS 232C offering the following features a Memory for up to 3900 values b Block segmentation up to 999 blocks Block No 1 Indication b001 Block No 2 Indication b002...

Страница 7: ...the Memory Mode It is possible to store the measuring values in different blocks in order to distinguish the different tasks To set up a new block e g Block 2 push both slides simultaneously forward...

Страница 8: ...o get any statistical value The display will then read AE Then the gauge automatically returns to the memory mode Deleting Measured Values While operating in the Memory Mode the QNix 1500 offers the p...

Страница 9: ...change Information the measured values will remain stored 6 bloc Displayed alternating with the block number 7 End Memory or block capacity out of space 8 N a alternating with Fe or NFe number of meas...

Страница 10: ...10 0 99 9 mil 1 mil in the range of 100 200 mil Repetitive accuracy 1 m 2 of reading in the range of 0 999 m 3 5 of reading in the range of 1000 5000 m Minimum object size 10 x 10 mm2 or 0 4 x 0 4 Mi...

Страница 11: ...11...

Страница 12: ...03 09 12 Automation Dr Nix GmbH Co KG Robert Perthel Str 2 50739 K ln Germany Tel 49 0 221 91 74 55 0 Fax 49 0 221 17 12 21 info automation de www automation de...

Отзывы: