nOTE: CHARGInG RATES VARY
AnD DEpEnD On THE:
The battery capacity rating (in minutes):
The higher the rate... the longer the charging
time required.
The degree of discharge:
A fully discharged battery requires 2 x’s as
much charging as 1/2 charged battery.
The battery condition and age:
An old battery and/or one in poor condition
requires more charging time.
The electrolytic temperature:
The colder the temperature...the longer the
charging time required.
CAuTIOn: Exceeding recommended
charging rates can damage battery plates
and generate potentially explosive gases.
APPENDIX I
BATTERY CHARGING GUIDE
APPENDIX II
TROUBLESHOOTING TESTER TROUBLE
Recommended charging rates for
fully discharged
batteries.
Battery Reserve
Charge Battery At:
Capacity (min.)
(slow charge on top
fast charge on bottom)
80 minutes or less
14 hrs. at 5 amps
7 hrs. at 10 amps
81-125 minutes
20 hrs. at 5 amps
10 hrs. at 10 amps
126-170 minutes
28 hrs. at 5 amps
14 hrs. at 10 amps
171-250 minutes
42 hrs. at 5 amps
21 hrs. at 10 amps
251 minutes and above
33 hrs. at 10 amps
PROBLEM
S
parks when attaching clamps
to battery
Carbon pile steams
Tester loads but has no volt
reading
Tester has volt reading but
does not load
Inaccurate amps reading
Incorrect ammeter polarity
Incorrect voltmeter polarity
SOLUTIONS
A. Turn load off
B. Turn power switch off
A. Not a problem-will remedy itself
B. Follow duty cycle closely (15 sec.
0n 45 sec. off)
A. Clean battery posts
B. Reposition clamp, both sides of
jaw must make contact
A. Clean battery posts
B. Reposition clamp, both sides of
jaw must make contact
A. Clean jaws
B. Adjust zero knob
A. Turn Inductive Amp Probe around
A. Switch volt leads connections
POSSIBLE CAUSES
A. Load turned on
B. Tester power switch on
A. Moisture in carbon discs
B. Overheating carbon pile
A. Poor connection on copper side of
clamp jaws.
A. Poor connection on silver side od
clamp jaws
A. Dirty Inductive Amp Probe jaws
B. Off zero
A. Inductive Amp Probe wrong direction
A. External volt leads connections
reversed