Auria Headpiece or
HiRes Auria Processor Gets Wet
Remove the Auria PowerCel immediately. Drain as much excess water as possible
from the processor or headpiece.
Place the processor and/or Auria Headpiece in the Auria Carrying Case immediately.
Contact your cochlear implant center or Advanced Bionics for further instructions.
NOTE: If your system appears to be working but you experience deterioration in
the volume or the quality and clarity of sound, contact your implant center
audiologist as your processor may need to be reprogrammed.
IMPORTANT: If exposure to fluids is suspected, attempt to dry processor in a warm
room for 24 hours. DO NOT try to speed up the drying process with
the use of a microwave oven, convection oven or hairdryer; as these
may further damage the internal electronics.
WARNING: Your
processor
or
headpiece
should
be
serviced
only
at
Advanced
Bionics. Do not attempt to open or repair the processor. Do not continue
to
use
the
HiRes
Auria
Processor
if
any
part
of
it
is
damaged.
Unauthorized opening of the processor, or other equipment, will void the
warranty and may compromise system performance.
Red Light-Emitting Diode (LED) on the Charger
When the red LED at a battery slot lights up, it indicates that the PowerCel should
be replaced.
Contact your cochlear implant center or Advanced Bionics for further instructions.
No Green LED on the Charger
•
Verify charger is connected to power.
•
Replace power supply, adapters, and/or charger.