33
*For all EQUIPMENT and SYSTEMS
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The customer
and/or user of this device should ensure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test
level
Compliance level
Electromagnetic
environment -guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete, or ceramic
tile. If floors are
covered with synthetic
material, relative
humidity should be at
least 30%.
Electrical Fast
Transient /
Burst
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power supply
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV Line to Line
±1 kV Line to Line
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0% UT in 0.5 cycle at
0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% UT in 1 cycle at 0°
70% UT in 25/30 cycles
at 0°
0% UT in 250/300
cycles at 0° and 180°
0% UT in 0.5 cycle at
0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% UT in 1 cycle at 0°
70% UT in 25/30 cycles
at 0°
0% UT in 250/300
cycles at 0° and 180°
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE: UT is the AC mains voltage before application of the test level.
Содержание Guardian Angel GA1000 Series
Страница 4: ...3 Guide to Symbols Single Use...
Страница 5: ...4 Date of Manufacture...
Страница 16: ...15 the button to clear the alarm of bluetooth disconnected or power off...