4 - INTERNAL FUNCTIONAL DIAGRAM
The first diagram shows us the unit set for
impedance
measurements (in
Internal
Mode
referring to CLIO software, see chapter 13 of the User's Manual); in this
situation the power amplifier and inputs are disconnected while the speaker load is
presented in parallel with CLIO's input and output; this method of impedance
measurement relies on the precise knowledge of the analyzer's output impedance and
stimulates the D.U.T with signals of small amplitude.
USB
To CLIO
Input 1
Input 2
Speaker
From CLIO
26 dB
I Sense
Input 4
Input 3
Pedal
IN OUT
5
6
Internal connections for impedance measurements (Internal Mode)
The second diagram shows us the unit set for
frequency response
measurements;
in this situation CLIO's output is fed to the power amplifier that drives the speaker
under test while one among four inputs is directed to CLIO's input.
To CLIO
Speaker
From CLIO
26 dB
I Sense
USB
Pedal
IN OUT
5
6
Input 1
Input 2
Input 4
Input 3
Internal connections for frequency response measurements
NOTE
: when the unit is set for frequency response measurements it is possible to
measure the current flowing in the D.U.T. through the
ISENSE
output; this permits
impedance measurements stimulating the D.U.T. with signals of variable amplitude up
to the amplifier's maximum power or voltage. Refer to chapter 13 of the CLIO User's
Manual.